Dural tissue trauma and cerebrospinal fluid leak after epidural needle puncture: effect of needle design, angle, and bevel orientation.

Pamela J Angle1, Jean E Kronberg, Dorothy E Thompson, Cameron Ackerley, John Paul Szalai, James Duffin, Peter Faure.   

Abstract

BACKGROUND: The effects of epidural needle design, angle, and bevel orientation on cerebrospinal fluid leak after puncture have not been reported. The impact of these factors on leak rate was examined using a dural sac model. Dural trauma was examined using scanning electron microscopy.
METHODS: Human cadaveric dura, mounted on a cylindrical model, was punctured with epidural needles using a micromanipulator. Tissue was punctured at 15 cm H2O (left lateral decubitus) system pressure, and leak was measured at 25 cm H2O (semisitting) pressure. Leak rates and trauma were compared for the following: (1) six different epidural needles at 90 degrees, bevel parallel to the dural long axis; (2) 18-gauge Tuohy and 18-gauge Special Sprotte epidural needles, 30 degrees versus 90 degrees; (3) 18-gauge Tuohy, bevel perpendicular versus parallel to the dural long axis.
RESULTS: With the 90 degrees puncture, bevel parallel, the greatest leak occurred with a 17-gauge Hustead (516 +/- 319 ml/15 min), and the smallest leak occurred with a 20-gauge Tuohy (100 +/- 112 ml/15 min; P = 0.0018). A 20-gauge Tuohy puncture led to statistically significant reductions in leak (P value range, 0.0001-0.0024) compared with all needles except the Special Sprotte. With the 30 degrees versus 90 degrees angle, 30 degrees punctures with an 18-gauge Tuohy produced nonstatistically significant leak reductions compared with the 18-gauge Tuohy at 90 degrees. The puncture angle made no difference for the Special Sprotte. Nonsignificant reductions were found for the Special Sprotte compared with the Tuohy. With the 18-gauge Tuohy bevel orientation, perpendicular orientation produced nonstatistically significant reductions in leak compared with parallel orientation.
CONCLUSIONS: Cerebrospinal fluid leak after puncture was influenced most by epidural needle gauge. Leak rate was significantly less for the 20-gauge Tuohy needle.

Entities:  

Mesh:

Year:  2003        PMID: 14639152     DOI: 10.1097/00000542-200312000-00021

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  6 in total

1.  Post spinal puncture headache, an old problem and new concepts: review of articles about predisposing factors.

Authors:  Ali Jabbari; Ebrahim Alijanpour; Mehrafza Mir; Nadia Bani Hashem; Seyed Mozaffar Rabiea; Mohammad Ali Rupani
Journal:  Caspian J Intern Med       Date:  2013

Review 2.  Needle gauge and tip designs for preventing post-dural puncture headache (PDPH).

Authors:  Ingrid Arevalo-Rodriguez; Luis Muñoz; Natalia Godoy-Casasbuenas; Agustín Ciapponi; Jimmy J Arevalo; Sabine Boogaard; Marta Roqué I Figuls
Journal:  Cochrane Database Syst Rev       Date:  2017-04-07

3.  Accidental dural puncture, postdural puncture headache, intrathecal catheters, and epidural blood patch: revisiting the old nemesis.

Authors:  Roland Kaddoum; Faisal Motlani; Romeo N Kaddoum; Arvi Srirajakalidindi; Deepak Gupta; Vitaly Soskin
Journal:  J Anesth       Date:  2013-12-18       Impact factor: 2.078

4.  Assessment of risk factors for postdural puncture headache in women undergoing cesarean delivery in Jordan: a retrospective analytical study.

Authors:  Wail N Khraise; Mohammed Z Allouh; Khaled M El-Radaideh; Raed S Said; Anas M Al-Rusan
Journal:  Local Reg Anesth       Date:  2017-03-17

5.  Introduction of Sprotte needles to a single-centre acute neurology service: before and after study.

Authors:  Vejay N Vakharia; Hazel Lote
Journal:  JRSM Short Rep       Date:  2012-12-15

6.  A Randomized Comparative Study on Median and Paramedian Approaches for Subarachnoid Block using Sprotte Needle in the Cesarean Section: Quest for the Best.

Authors:  Reena Kumari; Tushar Kumar; Boniface Hembrom; Barun Ram; Ladhu Lakra; Usha Suwalka
Journal:  Anesth Essays Res       Date:  2022-02-07