Sukhendu B Dev1, Deepak Dhar, Wanda Krassowska.
Abstract
We present an analytical solution for the electrical potential and field established by a six-needle array electroporation electrode, which is used in vivo for cancer treatment and DNA delivery. The analytical solution closely matches the numerical solution obtained with the finite element method: the mean error is less than 0.6%.Entities:
Mesh:
Substances:
Year: 2003 PMID: 14620000 DOI: 10.1109/TBME.2003.818467