On the field-size dependence of relative output from a linear accelerator.

P B Dunscombe1, J M Nieminen.   

Abstract

The radiation output in air on the central axis of a linac photon beam has been modeled as the sum of two components. These are a point source representing radiation direct from the target and a distributed source representing scatter in the flattening filter and primary collimator. By fitting only two parameters, the ratio of the two components for a 20 x 20 field and a width parameter for the distributed source this semi-empirical model describes the relative outputs of square, symmetric rectangular, and asymmetric rectangular fields with an average error of 0.25% for the field sizes studied.

Mesh:

Year:  1992        PMID: 1461207     DOI: 10.1118/1.596799

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Experimental evaluations of head scatter factor calculation by use of a Gaussian function.

Authors:  Ryoichi Notake; Yukishige Kyuuma; Kouji Sasamori; Makoto Sagawa; Yasushi Annaka; Hidetoshi Saitoh; Hitoshi Shibuya
Journal:  Radiol Phys Technol       Date:  2013-12-10

2.  Monitor unit calculations for external photon and electron beams: Report of the AAPM Therapy Physics Committee Task Group No. 71.

Authors:  John P Gibbons; John A Antolak; David S Followill; M Saiful Huq; Eric E Klein; Kwok L Lam; Jatinder R Palta; Donald M Roback; Mark Reid; Faiz M Khan
Journal:  Med Phys       Date:  2014-03       Impact factor: 4.071

3.  Prediction of back-scatter radiations to a beam monitor chamber of medical linear accelerators by use of the digitized target-current-pulse analysis method.

Authors:  Yusuke Suzuki; Naoki Hayashi; Hideki Kato; Hiroshi Fukuma; Yasujiro Hirose; Makoto Kawano; Yoshio Nishii; Masaru Nakamura; Takashi Mukouyama
Journal:  Radiol Phys Technol       Date:  2012-10-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.