G C Hurst1, J Hua, O P Simonetti, J L Duerk.
Abstract
This report describes NMR image effects due to sampling asymmetry when using zero-padded magnitude FT reconstruction. With this method, the MTF is not flat over the spatial frequency passband, so resolution cannot be accurately described by a single variable such as voxel size. At small to moderate asymmetry, shortened (reduced window duration) asymmetry provides increased S/N and decreased resolution, whereas shifted (constant window duration) asymmetry yields essentially constant S/N with simultaneously increased and decreased resolution. A bias function expression describes image distortion due to sampling in terms separable from the imaged object. The analyses are consistent with previous descriptions of perceived image differences related to data asymmetry.Entities:
Mesh:
Year: 1992 PMID: 1461112 DOI: 10.1002/mrm.1910270206
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668