[Experimental investigations at the new digital mammographic system].

R Schulz-Wendtland1, M Lell, E Wenkel, U Aichinger, K Imhoff, W Bautz.   

Abstract

PURPOSE: Experimental investigations of a new digital mammography system.
MATERIALS AND METHODS: The digital mammography system SenoScan (Fischer Imaging) is based on 4 charged coupled devices (CCD) (slot detector). The pixel size is 50 micro m, field of view 21 x 29 cm (2), effective quantum efficiency (DQE) (0) 50 %, spatial resolution 10 lp/mm, and memory depth 12 bit. The object of investigation was the Wisconsin Mammographic Random Phantom, Model 152 A (Radiation Measurements Inc., Wisconsin). The radiograms were acquired using the digital mammography system.
RESULTS: The digital mammography detected 225 of 225 possible details. The surface dose was 6.95 mGy.
CONCLUSIONS: These experimental results support the use of the digital mammography in practice. This has to be confirmed in further clinical trials.

Entities:  

Mesh:

Year:  2003        PMID: 14610710     DOI: 10.1055/s-2003-43407

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010