Literature DB >> 14606989

Validation of the BIACORE 3000 platform for detection of antibodies against erythropoietic agents in human serum samples.

Stephanie Mason1, Stephon La, Daniel Mytych, Steven J Swanson, John Ferbas.   

Abstract

OBJECTIVE: To develop a validated BIACORE immunoassay for the detection and characterization of serum antibodies with specificity for erythropoietic molecules (e.g. darbepoetin alfa).
METHODS: New Zealand White rabbits (n = 8) were immunized by an intramuscular injection of darbepoetin alfa/adjuvant at 0, 4, 6, and 8 weeks. Serum was collected for 6 weeks after final injection and pooled for affinity purification. Antibody immunoassay measurements were performed using a BIACORE 3000 with darbepoetin alfa immobilized to the biosensor surface. Human serum samples were spiked with the affinity-purified rabbit antibody to develop and validate the BIACORE immunoassay.
RESULTS: The assay was shown to be stable through 180 sample/regeneration cycles and had a threshold of 45.8 response units. The validated limit of detection was 0.40 microg/ml in 100% human serum. The method was robust, with variability not exceeding a 20% coefficient of variation, well within acceptable limits for typical immunoassays.
CONCLUSION: All protein-based therapeutics have a potential for immunogenicity, and antibodies raised against these molecules may have important clinical sequelae. The biotechnology and pharmaceutical industries are challenged to address this potential by developing robust analytical platforms to detect and characterize antibodies directed against therapeutic proteins in clinical specimens. Traditionally, radioimmune precipitation assays and/or enzyme-linked immunoassays (ELISAs) are used for primary detection of host immune response; however, the BIACORE platform may be better suited for this purpose in many instances. This platform represents a robust tool that should be considered for the detection and characterization of antibodies directed against protein-based therapeutics.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14606989     DOI: 10.1185/030079903125002414

Source DB:  PubMed          Journal:  Curr Med Res Opin        ISSN: 0300-7995            Impact factor:   2.580


  6 in total

1.  Directional surface plasmon-coupled emission: application for an immunoassay in whole blood.

Authors:  Evgenia G Matveeva; Zygmunt Gryczynski; Joanna Malicka; Joanna Lukomska; Slawomir Makowiec; Klaus W Berndt; Joseph R Lakowicz; Ignacy Gryczynski
Journal:  Anal Biochem       Date:  2005-09-15       Impact factor: 3.365

2.  An open-label extension study evaluating the safety and efficacy of romiplostim for up to 3.5 years in thrombocytopenic Japanese patients with immune thrombocytopenic purpura (ITP).

Authors:  Yukari Shirasugi; Kiyoshi Ando; Koji Miyazaki; Yoshiaki Tomiyama; Koji Iwato; Shinichiro Okamoto; Mineo Kurokawa; Keita Kirito; Satoshi Hashino; Haruhiko Ninomiya; Shinichiro Mori; Yuji Yonemura; Kensuke Usuki; Helen Wei; Richard Lizambri
Journal:  Int J Hematol       Date:  2012-04-25       Impact factor: 2.490

3.  Feasibility of a multiplex flow cytometric bead immunoassay for detection of anti-epoetin alfa antibodies.

Authors:  John Ferbas; John Thomas; John Hodgson; Amitabh Gaur; Nicole Casadevall; Steven J Swanson
Journal:  Clin Vaccine Immunol       Date:  2007-07-18

4.  Data quality in drug discovery: the role of analytical performance in ligand binding assays.

Authors:  Hermann Wätzig; Imke Oltmann-Norden; Franziska Steinicke; Hassan A Alhazmi; Markus Nachbar; Deia Abd El-Hady; Hassan M Albishri; Knut Baumann; Thomas Exner; Frank M Böckler; Sami El Deeb
Journal:  J Comput Aided Mol Des       Date:  2015-06-13       Impact factor: 3.686

5.  Anti-erythropoietin antibodies followed by endogenous erythropoietin production in a dialysis patient.

Authors:  Parnell Mattison; Kiran Upadhyay; Jennifer E Wilcox; Asha Moudgil; Douglas M Silverstein
Journal:  Pediatr Nephrol       Date:  2010-01-21       Impact factor: 3.714

Review 6.  Toward Exosome-Based Neuronal Diagnostic Devices.

Authors:  Yong Kyoung Yoo; Junwoo Lee; Hyungsuk Kim; Kyo Seon Hwang; Dae Sung Yoon; Jeong Hoon Lee
Journal:  Micromachines (Basel)       Date:  2018-11-29       Impact factor: 2.891

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.