| Literature DB >> 14604419 |
Abstract
The dosimetric characteristics for modern computer-controlled linear accelerators with the same make, model, and nominal energy are known to be very similar, as long as the machines are unaltered from the manufacturer's original specifications. In this preliminary study, a quantitative investigation of the similarity in the basic photon dosimetry data from the Siemens Primus linear accelerators at eight different institutions is reported. The output factor, percentage depth dose (PDD), and in-air off-axis factor (OAF) for the 6 and 18 MV photon beams measured or verified by the Radiological Physics Center (RPC) were analyzed. The RPC-measured output factors varied by less than about 2% for each field size. The difference between the maximum and minimum RPC-verified PDD values at each depth was less than about 3%. The difference between the maximum and minimum RPC-measured in-air OAF was no more than 4% at all off-axis distances considered in this study. These results strongly suggest that it is feasible to establish a reference photon dosimetry data set for each make, model, and nominal energy, universally applicable to those machines unaltered from the manufacturers' original specifications, within a clinically acceptable tolerance (e.g., approximately +/-2%). (c) 2003 American College of Medical Physics.Entities:
Mesh:
Year: 2003 PMID: 14604419 PMCID: PMC5724467 DOI: 10.1120/jacmp.v4i4.2500
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
The RPC‐measured output factors for the 6 MV photon beam from Siemens Primus accelerators. Note the averages were obtained from eight sets of measured data.
| Field size | Average OF | Max/min | %SD |
|---|---|---|---|
|
| 0.954 | 1.015 | 0.48 |
|
| 1.000 | ‐ | ‐ |
|
| 1.034 | 1.012 | 0.41 |
|
| 1.053 | 1.014 | 0.45 |
|
| 1.075 | 1.020 | 0.65 |
The RPC‐measured output factors for the 18 MV photon beam from Siemens Primus accelerators. Note the averages were obtained from eight sets of measured data.
| Field size | Average of | Max/min | %SD |
|---|---|---|---|
|
| 0.952 | 1.006 | 0.22 |
|
| 1.000 | ‐ | ‐ |
|
| 1.036 | 1.021 | 0.69 |
|
| 1.059 | 1.018 | 0.59 |
|
| 1.080 | 1.015 | 0.48 |
Makes and models of the chambers used.
| Ion chamber | Wall material | Cavity radius (cm) |
|---|---|---|
| NEL 2571 | Graphite | 0.315 |
| PTW N23333 | PMMA | 0.305 |
| PTW N30001 | PMMA | 0.305 |
| Exradin A‐12 | C‐552 | 0.305 |
PDD table for the 6 MV photon beam from Siemens Primus accelerators. Note: (i) the averages were obtained from 8 sets of the RPC‐verified institution data; (ii) max/min values for each average in this table are less than 1.03.
|
|
|
|
| |||||
|---|---|---|---|---|---|---|---|---|
| Depth (cm) | PDD | %SD | PDD | %SD | PDD | %SD | PDD | %SD |
| 1.5 | 100.0 | ‐ | 100.0 | ‐ | 100.0 | ‐ | 100.0 | ‐ |
| 2 | 99.3 | 0.30 | 99.1 | 0.29 | 99.1 | 0.21 | 98.8 | 0.26 |
| 3 | 95.0 | 0.24 | 95.2 | 0.42 | 95.5 | 0.32 | 95.5 | 0.37 |
| 4 | 90.3 | 0.43 | 91.1 | 0.45 | 91.7 | 0.34 | 91.8 | 0.48 |
| 5 | 85.8 | 0.34 | 86.9 | 0.47 | 88.3 | 0.42 | 88.5 | 0.41 |
| 6 | 81.2 | 0.21 | 82.7 | 0.50 | 84.3 | 0.39 | 85.1 | 0.40 |
| 7 | 76.8 | 0.24 | 78.7 | 0.38 | 80.7 | 0.49 | 81.6 | 0.48 |
| 8 | 72.5 | 0.42 | 74.6 | 0.34 | 77.1 | 0.52 | 78.3 | 0.46 |
| 9 | 68.4 | 0.54 | 70.9 | 0.49 | 73.6 | 0.59 | 74.8 | 0.79 |
| 10 | 64.8 | 0.45 | 67.3 | 0.61 | 70.5 | 0.56 | 71.8 | 0.68 |
| 15 | 48.3 | 0.72 | 51.3 | 0.69 | 55.4 | 0.76 | 57.4 | 0.66 |
| 20 | 36.1 | 0.84 | 38.9 | 0.93 | 43.2 | 0.96 | 45.3 | 0.78 |
| 25 | 27.0 | 0.91 | 29.3 | 0.67 | 33.5 | 0.60 | 35.5 | 0.95 |
| 30 | 20.4 | 0.85 | 22.4 | 1.17 | 26.0 | 1.16 | 28.0 | 0.55 |
PDD table for the 18 MV photon beam from Siemens Primus accelerators. Note: (i) the averages were obtained from eight sets of the RPC‐verified institution data; (ii) Note max/min values for each average in this table are less than 1.03; (iii) Atypical at a field size for 18 MV photon beam is 3.5 cm.
|
|
|
|
| |||||
|---|---|---|---|---|---|---|---|---|
| Depth (cm) | PDD | %SD | PDD | %SD | PDD | %SD | PDD | %SD |
|
| 100.0 | ‐ | 100.0 | ‐ | 100.0 | ‐ | 100.0 | ‐ |
| 4 | 98.9 | 0.23 | 98.8 | 0.22 | 98.3 | 0.34 | 97.5 | 0.95 |
| 5 | 95.8 | 0.49 | 95.7 | 0.42 | 94.9 | 0.52 | 94.6 | 0.49 |
| 6 | 91.8 | 0.50 | 92.0 | 0.39 | 91.7 | 0.53 | 91.2 | 0.81 |
| 7 | 88.0 | 0.48 | 88.3 | 0.45 | 88.3 | 0.47 | 88.0 | 0.75 |
| 8 | 84.3 | 0.57 | 84.9 | 0.46 | 85.1 | 0.49 | 84.9 | 0.73 |
| 9 | 80.7 | 0.55 | 81.5 | 0.44 | 81.9 | 0.39 | 81.9 | 0.72 |
| 10 | 77.3 | 0.54 | 78.1 | 0.39 | 78.8 | 0.58 | 79.1 | 0.72 |
| 15 | 61.9 | 0.65 | 63.4 | 0.51 | 64.9 | 0.61 | 65.6 | 0.59 |
| 20 | 49.6 | 0.67 | 51.2 | 0.57 | 53.4 | 0.80 | 54.4 | 0.49 |
| 25 | 39.8 | 0.96 | 41.5 | 0.92 | 43.8 | 0.83 | 44.8 | 1.11 |
| 30 | 32.1 | 1.12 | 33.7 | 1.12 | 35.9 | 1.38 | 37.1 | 0.61 |
In‐air off‐axis factors for the 6 MV photon beam from Siemens Primus accelerators. Note: (i) The averages were obtained from 8 sets of the RPC‐measured data; (ii) The presented values may be different from in‐water OAF.
| Off‐axis distance (cm) | Average OAF | Max/min | %SD |
|---|---|---|---|
| 5 | 1.042 | 1.014 | 0.42 |
| 10 | 1.058 | 1.024 | 0.86 |
| 15 | 1.075 | 1.037 | 1.30 |
In‐air off‐axis factors for the 18 MV photon beam from Siemens Primus accelerators. Note: (i) The averages were obtained from 8 sets of the RPC‐measured data; (ii) The presented values may be different from in‐water OAF.
| Off‐axis distance (cm) | Average OAF | Max/min | %SD |
|---|---|---|---|
| 5 | 1.054 | 1.033 | 1.01 |
| 10 | 1.060 | 1.035 | 1.12 |
| 15 | 1.063 | 1.040 | 1.32 |