What is the optimal paddle force during paediatric external defibrillation?

Charles D Deakin1, Sarah H Bennetts, Graham W Petley, Frank Clewlow.   

Abstract

INTRODUCTION: Transthoracic impedance (TTI) is a major determinant of transmyocardial current flow, and therefore, the success of defibrillation. European Resuscitation Council (ERC) paediatric guidelines recommend that 'firm' paddle force should be applied to the paddles during defibrillation. No study has yet established the optimal paddle force required to minimise TTI in children of different ages.
METHODS: Eighty patients aged 10 weeks to 17 yrs undergoing general anaesthesia for routine surgery were studied. Using defibrillation paddles placed in an anterior-apical position, TTI (Omega) was measured for increasing values of force from 0.5 kgf (baseline) to 6.5 kgf. The optimal force, the force to achieve 95% of the overall reduction in TTI, was then determined. According to current guidelines, paediatric paddles (surface area 16 cm2) were used for infants (< or =10 kg) and adult paddles (82 cm2) for older children. Optimal force was then calculated for infants < or =10 kg, children >10 kg and < or =8 yrs and children 9-17 yrs age.
RESULTS: Increasing paddle force from 0.5 kgf progressively decreased TTI. Optimal force using paediatric paddles was 2.9 kgf in infants. Optimal force using adult paddles was 5.1 kgf in children >10 kg but < or = 8 yrs and 5.3 kgf in children aged 9-17 yrs.
CONCLUSIONS: Force is an important determinant of TTI and therefore, outcome of defibrillation. It is recommended that a minimum of 3 kgf be applied to paddles when defibrillating infants with paediatric paddles, and a minimum of 5 kgf be applied to all older children when adult paddles are used.

Entities:  

Mesh:

Year:  2003        PMID: 14580737     DOI: 10.1016/s0300-9572(03)00173-4

Source DB:  PubMed          Journal:  Resuscitation        ISSN: 0300-9572            Impact factor:   5.262


  4 in total

1.  Analysis of transthoracic impedance during real cardiac arrest defibrillation attempts in older children and adolescents: are stacked-shocks appropriate?

Authors:  Dana E Niles; Akira Nishisaki; Robert M Sutton; Sara Brunner; Mette Stavland; Shruthi Mahadevaiah; Peter A Meaney; Matthew R Maltese; Robert A Berg; Vinay M Nadkarni
Journal:  Resuscitation       Date:  2010-08-13       Impact factor: 5.262

Review 2.  Ventricular fibrillation and defibrillation.

Authors:  P Jones; N Lodé
Journal:  Arch Dis Child       Date:  2007-10       Impact factor: 3.791

Review 3.  A Systematic Review of the Transthoracic Impedance during Cardiac Defibrillation.

Authors:  Yasmine Heyer; Daniela Baumgartner; Christian Baumgartner
Journal:  Sensors (Basel)       Date:  2022-04-06       Impact factor: 3.576

Review 4.  Electrical cardioversion.

Authors:  Murat Sucu; Vedat Davutoglu; Orhan Ozer
Journal:  Ann Saudi Med       Date:  2009 May-Jun       Impact factor: 1.526

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.