Continuous measurement of blood volume during hemodialysis by an optical method.

J P de Vries1, C G Olthof, V Visser, P M Kouw, A van Es, J M Donker, P M de Vries.   

Abstract

A new method is described to noninvasively and continuously measure changes in blood volume (BV) during hemodialysis by means of an optical reflection method with an optical monitor (950 nm) clipped onto the arterial blood line. The amount of reflected light (L) appeared to be linearly proportional to the erythrocyte concentration (r = 0.91). Changes in L correlated well with changes in erythrocyte concentration during hemodialysis (r = 0.94). A study in 10 patients on regular dialysis was done. The BV decrease after 3 hr of treatment was 17.0 +/- 5.2%, and it correlated with the amount of fluid withdrawn by ultrafiltration (mean, 2,519 +/- 589 ml). Five hypotensive episodes were seen that were characterized by a higher rate of BV fall during the preceding 15 min (9.9 +/- 1.9 versus 3.6 +/- 4.3%/hrp; p < 0.05) and by a lower BV value at that moment (78.2 +/- 3.4 versus 84.5 +/- 4.5%; p < 0.025) than in the other five patients at comparable times. It was concluded that this optical method was a means to detect hypovolemia at an early stage and to prevent ultrafiltration induced hypotension.

Entities:  

Mesh:

Year:  1992        PMID: 1457844     DOI: 10.1097/00002480-199207000-00015

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  5 in total

1.  Comparison of optical, electrical, and centrifugation techniques for haematocrit monitoring of dialysed patients.

Authors:  M Y Jaffrin; C Fournier
Journal:  Med Biol Eng Comput       Date:  1999-07       Impact factor: 2.602

2.  Continuous measurements by impedance of haematocrit and plasma volume variations during dialysis.

Authors:  M Maasrani; M Y Jaffrin; B Boudailliez
Journal:  Med Biol Eng Comput       Date:  1997-05       Impact factor: 2.602

3.  Comparison of markers of circulating blood volume in hemodialysis patients.

Authors:  Yoshiyuki Morishita; Yasuhiro Ando; Eriko Ishii; Mayumi Arisaka; Eiji Kusano
Journal:  Clin Exp Nephrol       Date:  2005-09       Impact factor: 2.801

4.  Continuous measurement of hematocrit using an intravascular catheter equipped with a fiberoptic transmission cell.

Authors:  Hiroko Fukushima
Journal:  J Anesth       Date:  1998-06       Impact factor: 2.078

5.  Cardiac output and circulating blood volume analysis by pulse dye-densitometry.

Authors:  T Iijima; T Aoyagi; Y Iwao; J Masuda; M Fuse; N Kobayashi; H Sankawa
Journal:  J Clin Monit       Date:  1997-03
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.