Literature DB >> 1451230

Inspiratory valve malfunction in a circle system: pitfalls in capnography.

A Y Kumar1, K Bhavani-Shankar, H S Moseley, Y Delph.   

Abstract

Capnography is a useful technique in monitoring the integrity of anaesthetic equipment such as the malfunctioning of unidirectional valves in circle system. However, the lack of a precise mechanism in existing capnographs to identify the start of inspiration and the beginning of expiration in the capnograms, makes the analysis of the carbon dioxide waveforms during inspiration difficult and thus results in inaccurate assessment of rebreathing. We report a case where, during the malfunction of the inspiratory unidirectional valve in the circle system, the capnograph failed to detect the presence of substantial rebreathing. Critical analysis of the capnogram recorded during the malfunction revealed that there was substantial rebreathing which was underestimated by the capnograph as it reports only the lowest CO2 concentration rebreathed during inspiration in such abnormal situations.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1451230     DOI: 10.1007/BF03008353

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  4 in total

1.  Respiratory monitoring of carbon dioxide and oxygen: a ten-year perspective.

Authors:  M Weingarten
Journal:  J Clin Monit       Date:  1990-07

2.  Capnographic detection of anaesthesia circle valve malfunctions.

Authors:  L S Berman; S T Pyles
Journal:  Can J Anaesth       Date:  1988-09       Impact factor: 5.063

3.  Expiratory valve dysfunction in a semiclosed circle anesthesia circuit--verification by analysis of carbon dioxide waveform.

Authors:  S T Pyles; L S Berman; J H Modell
Journal:  Anesth Analg       Date:  1984-05       Impact factor: 5.108

4.  Computer-assisted capnogram analysis.

Authors:  H R van Genderingen; N Gravenstein; J J van der Aa; J S Gravenstein
Journal:  J Clin Monit       Date:  1987-07
  4 in total
  1 in total

Review 1.  Terminology and the current limitations of time capnography: a brief review.

Authors:  K Bhavani-Shankar; A Y Kumar; H S Moseley; R Ahyee-Hallsworth
Journal:  J Clin Monit       Date:  1995-05
  1 in total