Literature DB >> 1424128

Measurement of cyclosporine by liquid chromatography and three immunoassays in blood from liver, cardiac, and renal transplant recipients.

J H McBride1, S S Kim, D O Rodgerson, A F Reyes, M K Ota.   

Abstract

In an effort to replace HPLC for whole-blood determination of cyclosporine (CsA), we compared HPLC with radioimmunoassay (RIA; INCSTAR, Cyclo-Trac SP assay), fluorescence polarization immunoassay (FPIA; Abbott TDx), and in-house modified enzyme-multiplied immunoassay technique (EMIT; Syva Co.). For blood samples from 200 various transplant recipients, RIA = 1.262 (HPLC) - 8.16, r = 0.983; FPIA = 1.200 (HPLC) + 19.90, r = 0.981; and EMIT = 1.038 (HPLC) + 11.28, r = 0.985. For segregation by transplant type, RIA, FPIA, and EMIT demonstrated positive biases of 27%, 12%, and 3%, respectively, for liver transplant recipients (n = 50) when compared with HPLC. Heart transplant recipients (n = 50) gave positive bias values of 23%, 14%, and 4% for RIA, FPIA, and EMIT, respectively. Adult renal transplant recipients (n = 50) demonstrated positive bias values of 30%, 31%, and 0% for RIA, FPIA, and EMIT, respectively. For pediatric renal transplant recipients (n = 50), positive biases of 40%, 31%, and 9% were obtained for RIA, FPIA, and EMIT, respectively. We conclude that the modified EMIT represents the best replacement for HPLC.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1424128

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  4 in total

Review 1.  The use of therapeutic drug monitoring to optimise immunosuppressive therapy.

Authors:  S M Tsunoda; F T Aweeka
Journal:  Clin Pharmacokinet       Date:  1996-02       Impact factor: 6.447

2.  Conversion of cardiac and liver transplant recipients from HPLC and FPIA (polyclonal) to an FPIA (monoclonal) technique for measurement of blood cyclosporin A.

Authors:  J H McBride; S Kim; D O Rodgerson; A Reyes
Journal:  J Clin Lab Anal       Date:  1998       Impact factor: 2.352

3.  Multisite analytical evaluation of the Abbott ARCHITECT cyclosporine assay.

Authors:  Pierre Wallemacq; Gregory T Maine; Keith Berg; Thomas Rosiere; Pierre Marquet; Giuseppe Aimo; Giulio Mengozzi; Julianna Young; Kurt Wonigeit; Robert Kretschmer; Bendicht Wermuth; Rainer W Schmid
Journal:  Ther Drug Monit       Date:  2010-04       Impact factor: 3.681

Review 4.  Cyclosporin. A review of its pharmacodynamic and pharmacokinetic properties, and therapeutic use in immunoregulatory disorders.

Authors:  Diana Faulds; Karen L Goa; Paul Benfield
Journal:  Drugs       Date:  1993-06       Impact factor: 9.546

  4 in total