Comparison of computerized digital and film-screen radiography: response to variation in imaging kVp.

N J Broderick1, B Long, R G Dreesen, M D Cohen, D A Cory, B P Katz, L A Kalasinski.   

Abstract

A controlled prospective study, in an animal model chosen to simulate portable neonatal radiography, was performed to compare the response of the Philips Computed Radiography (CR) system and conventional 200 speed film-screen (FS) to variation in imaging kVp. Acceptable images were obtained on the CR system over a very wide kVp range. In contrast the FS system produced acceptable images over a narrow kVp range. This ability suggests that the CR system should eliminate the need for repeat examinations in cases where a suboptimal kVp setting would have resulted in an unacceptable FS image. CR technology should therefore be ideally suited to portable radiography especially in situations where selection of correct exposure factors is difficult as in the neonatal nursery.

Entities:  

Mesh:

Year:  1992        PMID: 1408442     DOI: 10.1007/bf02016253

Source DB:  PubMed          Journal:  Pediatr Radiol        ISSN: 0301-0449


  7 in total

Review 1.  Digital imaging of the chest.

Authors:  R G Fraser; C Sanders; G T Barnes; H MacMahon; M L Giger; K Doi; A W Templeton; G G Cox; S J Dwyer; C R Merritt
Journal:  Radiology       Date:  1989-05       Impact factor: 11.105

2.  Use of computed radiography in respiratory distress syndrome in the neonatal nursery.

Authors:  Y Nakano; K Odagiri
Journal:  Pediatr Radiol       Date:  1989

3.  Excretory urography using computed radiography.

Authors:  L L Fajardo; B J Hillman; T B Hunter; H R Claypool; B R Westerman; B Mockbee
Journal:  Radiology       Date:  1987-02       Impact factor: 11.105

4.  Computed radiographic imaging: use in low-dose leg length radiography.

Authors:  M S Kogutt
Journal:  AJR Am J Roentgenol       Date:  1987-06       Impact factor: 3.959

Review 5.  Digital radiography: a review.

Authors:  D J Kastan; L V Ackerman; P J Feczko; G H Beute
Journal:  Henry Ford Hosp Med J       Date:  1985

6.  Computed radiography utilizing scanning laser stimulated luminescence.

Authors:  M Sonoda; M Takano; J Miyahara; H Kato
Journal:  Radiology       Date:  1983-09       Impact factor: 11.105

7.  Low-dose digital computed radiography in pediatric chest imaging.

Authors:  M S Kogutt; J P Jones; D D Perkins
Journal:  AJR Am J Roentgenol       Date:  1988-10       Impact factor: 3.959

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.