Literature DB >> 1385506

Transcatheter umbrella closure of valvular and paravalvular leaks.

M Hourihan1, S B Perry, V S Mandell, J F Keane, J J Rome, J A Bittl, J E Lock.   

Abstract

OBJECTIVES: Our aim was to adapt the technique of transcatheter umbrella closure of intracardiac defects for closure of valvular and paravalvular defects.
BACKGROUND: The double-umbrella device developed by Rashkind and Cuaso has been safely and effectively delivered across a host of intracardiac defects, but transcatheter closure of valvular and paravalvular leaks has not been reported.
METHODS: Between February 1987 and September 1990, eight patients who were believed to be poor operative candidates were taken to the catheterization laboratory for transcatheter double-umbrella closure of a valvular or a paravalvular leak. Four patients had a paravalvular leak around a prosthetic aortic valve. The other four patients had a valvular leak: one patient with a regurgitant native aortic valve after a Stansel procedure and three patients with a regurgitant porcine valve in a left ventricular apex to descending aorta conduit.
RESULTS: Placement of a double-umbrella device was attempted in seven of the eight patients and was successful in all seven. Device placement was not attempted in one patient because of the crescentic shape of his defect. Two patients required two devices for each closure; the other five required only one device each. Angiography, performed on six patients after device closure, demonstrated that three patients had a completely occluded defect, two had trivial residual flow and one patient had mild residual flow through the device. All significant complications occurred in one patient who had hemolysis and oliguria that resolved when the initial umbrella was replaced by a larger device. In addition, two devices migrated to the patient's pulmonary arteries but were retrieved in the catheterization laboratory without difficulty. No other early or late complications occurred in 21 to 50 months of follow-up. Of the four patients with a paravalvular leak, the one who did not receive a device died at operation, one patient died at operation for an associated defect (in the operating room the umbrella was found securely in place across the paraaortic defect) and two patients are clinically well at home after 21 and 32 months, respectively. Of the four patients with closure of a valvular leak, one patient remains well at home 50 months later, one patient died at operation for associated defects and two patients had additional successful surgical treatment and remain well 29 months after device placement.
CONCLUSIONS: Transcatheter umbrella closure appears to be a reasonable alternative for closure of a valvular or paravalvular leak in patients who are poor operative candidates.

Entities:  

Mesh:

Year:  1992        PMID: 1385506     DOI: 10.1016/0735-1097(92)90250-q

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  31 in total

1.  Real-time three-dimensional TEE guidance of a paravalvular mitral valve leak closure procedure.

Authors:  N Wunderlich; J Franke; H Sievert
Journal:  Herz       Date:  2011-11-17       Impact factor: 1.443

2.  Repair of paravalvular prosthetic mitral valve leaks with septal occluder devices in severely high-risk patients: a word of caution.

Authors:  Craig R Smith; Sotiris C Stamou; William M Merhi; Robert L Hooker
Journal:  Interact Cardiovasc Thorac Surg       Date:  2012-05-28

3.  Cisplatin-induced genes as potential markers for thyroid cancer.

Authors:  G Lapouge; R Millon; D Muller; J Abecassis; M Eber; J P Bergerat; C Klein-Soyer
Journal:  Cell Mol Life Sci       Date:  2005-01       Impact factor: 9.261

4.  Transcatheter closure of a ruptured sinus of Valsalva aneurysm (RSVA) after aortic valve replacement using the Amplatzer muscular VSD Occluder.

Authors:  R Schaeffler; S Sarikouch; M Peuster
Journal:  Clin Res Cardiol       Date:  2007-09-10       Impact factor: 5.460

Review 5.  Transcatheter approaches to non-valvar structural heart disease.

Authors:  Damien Kenny; Ziyad M Hijazi
Journal:  Int J Cardiovasc Imaging       Date:  2011-02-18       Impact factor: 2.357

Review 6.  Percutaneous mitral and aortic paravalvular leak repair: indications, current application, and future directions.

Authors:  Ronald K Binder; John G Webb
Journal:  Curr Cardiol Rep       Date:  2013-03       Impact factor: 2.931

7.  Interventional treatment of paravalvular regurgitation by plug implantation following prosthetic valve replacement: a single-center experience.

Authors:  Nicolas Werner; Uwe Zeymer; Boris Fraiture; Caroline Kilkowski; Patrick Riedmaier; Steffen Schneider; Ralf Zahn
Journal:  Clin Res Cardiol       Date:  2018-06-12       Impact factor: 5.460

Review 8.  Acetyl-coenzyme A synthetase (AMP forming).

Authors:  V J Starai; J C Escalante-Semerena
Journal:  Cell Mol Life Sci       Date:  2004-08       Impact factor: 9.261

9.  Interventional closure of two fistulas after aortic valve surgery.

Authors:  Robert Dalla Pozza; R Kozlik-Feldmann; T-P Le; C Schmitz; R Sodian; B Reichart; Heinrich Netz
Journal:  Clin Res Cardiol       Date:  2009-04-09       Impact factor: 5.460

10.  A case demonstrating a percutaneous closure using the amplatzer duct occluder for paravalvular leakage after tricuspid valve replacement.

Authors:  Young Hee Heo; Soo-Jin Kim; Sang Yun Lee; Jae Suk Baek
Journal:  Korean Circ J       Date:  2013-04-30       Impact factor: 3.243

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.