Literature DB >> 1342584

Signal delivery/real ear measurement system for hearing aid selection and fitting.

D L Zelisko1, R C Seewald, J P Gagné.   

Abstract

This report describes a signal delivery/real ear measurement system for application in the hearing aid selection and fitting process. This signal delivery/real ear measurement system provides a means for quantifying a listener's auditory characteristics (e.g., thresholds, loudness discomfort levels) in a manner that is compatible with electroacoustic measures of hearing aid performance. The signal delivery/real ear measurement system consists of a button-type hearing aid receiver coupled to the listener's personal earmold, and a probe tube microphone system to measure the signal level within the occluded ear canal. This signal delivery/real ear measurement system was used to measure detection thresholds and loudness discomfort levels in severe/profoundly hearing-impaired school-age children, with results indicating good test-retest reliability in behavioral responses. The findings presented in this report relate to the intersession electroacoustic variability associated with this instrumentation. In addition, the potential application of this or a similar system for measuring real ear to 2-cc coupler difference values is illustrated and discussed.

Entities:  

Mesh:

Year:  1992        PMID: 1342584

Source DB:  PubMed          Journal:  Ear Hear        ISSN: 0196-0202            Impact factor:   3.570


  2 in total

Review 1.  Clinical protocols for hearing instrument fitting in the Desired Sensation Level method.

Authors:  Marlene Bagatto; Sheila Moodie; Susan Scollie; Richard Seewald; Shane Moodie; John Pumford; K P Rachel Liu
Journal:  Trends Amplif       Date:  2005

2.  An approach for ensuring accuracy in pediatric hearing instrument fitting.

Authors:  Richard C Seewald; Susan D Scollie
Journal:  Trends Amplif       Date:  2003