Significance of peak value and duration of the interfacial shear load in evaluation of the bone-implant interface.

Yong-Shun Chang1, Masanori Kobayashi, Zhong-Lian Li, Masanori Oka, Takashi Nakamura.   

Abstract

OBJECTIVE: To determine if there is any merit in measuring peak load duration in studies of bone-implant interface during a pull-out test was performed to record the interfacial shear load.
DESIGN: Mechanical and histological investigations were performed to determine the bone-implant interfacial strength.
BACKGROUND: A common long-term problem following artificial joint replacement is the formation of an intervening fibrous tissue seam at the interface between bone and a prosthesis. To investigate the nature and strength of interfacial attachment between bone and an implanted biomaterial, a detaching method such as pull-out test can be used.
METHODS: Four porous surfaced alumina ceramic implants with or without glass-ceramic containing apatite and wollastonite coating at the surface of the beads and the substrate were implanted into canine the femoral condyles. Peak value and its duration of the interfacial shear load for the different types of implants were analyzed. Tissue formed at the bone-implant interfaces were examined histologically.Results. The peak value of interfacial shear load of the apatite and wollastonite glass-ceramic-coated implant was significantly greater than that of the non-coated implant at every stage, but the peak value duration in the control at each time period was greater than that of the coated implant. Histologically, more bone tissue ingrowth in the pores of coated implants, whereas a gap was observed at the interface of bone and the non-coated implant.
CONCLUSIONS: The interfacial shear load values, peak and its duration, are significantly influenced by the direct bone bonding or the intervening fibrous tissue which formed at the bone-implant interface. RELEVANCE: The peak load measurement and the duration of the peak load value are useful to measure the strength of interfacial attachment between bone and an implanted biomaterial. However, the duration of peak load value is closely related to the strength and plasticity of the tissue formed in the bone-implant interface. The analysis of peak value duration could be used to determine the properties of new grown tissue before histologic examination.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12957565     DOI: 10.1016/s0268-0033(03)00138-4

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  5 in total

Review 1.  Polyethylene and metal wear particles: characteristics and biological effects.

Authors:  Isabelle Catelas; Markus A Wimmer; Sandra Utzschneider
Journal:  Semin Immunopathol       Date:  2011-01-26       Impact factor: 9.623

2.  Emerging ideas: can erythromycin reduce the risk of aseptic loosening?

Authors:  Weiping Ren; David C Markel
Journal:  Clin Orthop Relat Res       Date:  2011-05-17       Impact factor: 4.176

3.  Cell-based osteoprotegerin therapy for debris-induced aseptic prosthetic loosening on a murine model.

Authors:  L Zhang; T-H Jia; A C M Chong; L Bai; H Yu; W Gong; P H Wooley; S-Y Yang
Journal:  Gene Ther       Date:  2010-04-29       Impact factor: 5.250

4.  The effect of adhesive strength of hydroxyapatite coating on the stability of hydroxyapatite-coated prostheses in vivo at the early stage of implantation.

Authors:  Yonghong Duan; Shu Zhu; Fei Guo; Jinyu Zhu; Mao Li; Jie Ma; Qingsheng Zhu
Journal:  Arch Med Sci       Date:  2012-05-09       Impact factor: 3.318

5.  Therapeutic potentials of naringin on polymethylmethacrylate induced osteoclastogenesis and osteolysis, in vitro and in vivo assessments.

Authors:  Nianhu Li; Zhanwang Xu; Paul H Wooley; Jianxin Zhang; Shang-You Yang
Journal:  Drug Des Devel Ther       Date:  2013-12-10       Impact factor: 4.162

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.