Diamond detector versus silicon diode and ion chamber in photon beams of different energy and field size.

M Bucciolini1, F Banci Buonamici, S Mazzocchi, C De Angelis, S Onori, G A P Cirrone.   

Abstract

The aim of this work was to test the suitability of a PTW diamond detector for nonreference condition dosimetry in photon beams of different energy (6 and 25 MV) and field size (from 2.6 cm x 2.6 cm to 10 cm x 10 cm). Diamond behavior was compared to that of a Scanditronix p-type silicon diode and a Scanditronix RK ionization chamber. Measurements included output factors (OF). percentage depth doses (PDD) and dose profiles. OFs measured with diamond detector agreed within 1% with those measured with diode and RK chamber. Only at 25 MV, for the smallest field size, RK chamber underestimated OFs due to averaging effects in a pointed shaped beam profile. Agreement was found between PDDs measured with diamond detector and RK chamber for both 6 MV and 25 MV photons and down to 5 cm x 5 cm field size. For the 2.6 cm x 2.6 cm field size, at 25 MV, RK chamber underestimated doses at shallow depth and the difference progressively went to zero in the distal region. PDD curves measured with silicon diode and diamond detector agreed well for the 25 MV beam at all the field sizes. Conversely, the nontissue equivalence of silicon led, for the 6 MV beam, to a slight overestimation of the diode doses in the distal region, at all the field sizes. Penumbra and field width measurements gave values in agreement for all the detectors but with a systematic overestimate by RK measurements. The results obtained confirm that ion chamber is not a suitable detector when high spatial resolution is required. On the other hand, the small differences in the studied parameters, between diamond and silicon systems, do not lead to a significant advantage in the use of diamond detector for routine clinical dosimetry.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12945980     DOI: 10.1118/1.1591431

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  Total scatter factors of small beams: a multidetector and Monte Carlo study.

Authors:  Paolo Francescon; Stefania Cora; Carlo Cavedon
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

2.  Assessment of the setup dependence of detector response functions for mega-voltage linear accelerators.

Authors:  Christopher Fox; Tom Simon; Bill Simon; James F Dempsey; Darren Kahler; Jatinder R Palta; Chihray Liu; Guanghua Yan
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

3.  Silicon diodes as an alternative to diamond detectors for depth dose curves and profile measurements of photon and electron radiation.

Authors:  Christian Scherf; Christiane Peter; Jussi Moog; Jörg Licher; Eugen Kara; Klemens Zink; Claus Rödel; Ulla Ramm
Journal:  Strahlenther Onkol       Date:  2009-08-04       Impact factor: 3.621

4.  Ionizing radiation detectors based on Ge-doped optical fibers inserted in resonant cavities.

Authors:  Saverio Avino; Vittoria D'Avino; Antonio Giorgini; Roberto Pacelli; Raffaele Liuzzi; Laura Cella; Paolo De Natale; Gianluca Gagliardi
Journal:  Sensors (Basel)       Date:  2015-02-12       Impact factor: 3.576

5.  Analysis of small field percent depth dose and profiles: Comparison of measurements with various detectors and effects of detector orientation with different jaw settings.

Authors:  Henry Finlay Godson; M Ravikumar; S Sathiyan; K M Ganesh; Y Retna Ponmalar; C Varatharaj
Journal:  J Med Phys       Date:  2016 Jan-Mar

6.  Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector.

Authors:  Suresh H Chaudhari; Rishabh Dobhal; Rajesh A Kinhikar; Sudarshan S Kadam; Deepak D Deshpande
Journal:  J Med Phys       Date:  2017 Jan-Mar

7.  Dose and slice thickness evaluation with nMAG gel dosimeters in computed tomography.

Authors:  Chun-Chao Chuang; Jay Wu
Journal:  Sci Rep       Date:  2018-02-08       Impact factor: 4.379

Review 8.  From cultured to uncultured genome sequences: metagenomics and modeling microbial ecosystems.

Authors:  Daniel R Garza; Bas E Dutilh
Journal:  Cell Mol Life Sci       Date:  2015-08-09       Impact factor: 9.261

9.  Radiographic film dosimetry for IMRT fields in the nearsurface buildup region.

Authors:  Peter L Roberson; Jean M Moran; Ravi Kulasekere
Journal:  J Appl Clin Med Phys       Date:  2008-10-24       Impact factor: 2.102

10.  The clinical impact of detector choice for beam scanning.

Authors:  Jacob A Gersh; Ryan C M Best; Ronald J Watts
Journal:  J Appl Clin Med Phys       Date:  2014-07-08       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.