Abstract

A retrospective study of the clinical and radiographic results of Mallory Head total hip arthroplasties in 67 patients (76 hips) with a mean age of 50.1 years was undertaken. The mean follow-up period was 10.1 years (9-13years). In this study, excellent or good clinical results were found for 64 patients. Minimal thigh pain was found in 3 patients (4.4%). Seventy-one hips (93.4%) showed fixation by bony ingrowth and 3 (3.9%) showed stable fibrous fixation. Bony ongrowth and ingrowth were also seen in 16 hips (21%) at distal smooth and grit-blast areas. Two femoral components were revised (survival rate, 97.3%), one for subsidence and other for recurrent infection. We believe that these good results were caused by excellent bony ingrowth on the proximal two thirds of the femoral surface area. In conclusion, proximal plasma-spray coating and the tapered shape of the Ti stem showed excellent bony ingrowth and initial stability. Our results indicate that the tapered shape will be an important element in the design of a cementless Ti stem.

Entities:  

Mesh:

Year:  2003        PMID: 12934216     DOI: 10.1016/s0883-5403(03)00203-1

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  5 in total

1.  Survivorship of 2000 tapered titanium porous plasma-sprayed femoral components.

Authors:  Adolph V Lombardi; Keith R Berend; Thomas H Mallory; Michael D Skeels; Joanne B Adams
Journal:  Clin Orthop Relat Res       Date:  2008-10-31       Impact factor: 4.176

2.  Total hip replacement arthroplasty with Mallory-Head system--minimum ten-year follow-up results.

Authors:  Joo-Hyoun Song; Dong-Hyun Kim; Juyoung Kim
Journal:  Int Orthop       Date:  2012-08-04       Impact factor: 3.075

3.  A short tapered stem reduces intraoperative complications in primary total hip arthroplasty.

Authors:  Ryan G Molli; Adolph V Lombardi; Keith R Berend; Joanne B Adams; Michael A Sneller
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

Review 4.  Implant Design in Cementless Hip Arthroplasty.

Authors:  Jung Taek Kim; Jeong Joon Yoo
Journal:  Hip Pelvis       Date:  2016-06-30

5.  Minimal stress shielding with a Mallory-Head titanium femoral stem with proximal porous coating in total hip arthroplasty.

Authors:  Brad Ellison; Nicholas A Cheney; Keith R Berend; Adolph V Lombardi; Thomas H Mallory
Journal:  J Orthop Surg Res       Date:  2009-12-09       Impact factor: 2.359