Portable spectral Doppler echocardiographic device: overcoming limitations.

J Quiles1, M A García-Fernández, P B Almeida, E Pérez-David, J Bermejo, M Moreno, P Avanzas.   

Abstract

BACKGROUND: There is evidence that new portable echocardiographic devices are useful in evaluating heart anatomy and function, but a lack of Doppler modes has up to now been an important limitation in obtaining haemodynamic data.
OBJECTIVES: To report the Doppler capabilities of a new hand held echocardiographic device.
DESIGN: Blinded comparison of two types of echocardiography machine.
SETTING: Tertiary care centre. PATIENTS: 98 consecutive patients were randomly imaged with the hand held device, with a standard platform as reference. OUTCOME MEASURES: Pulsed wave transmitral Doppler inflow tract velocities, deceleration time, and continuous wave Doppler measurements of aortic ejection and tricuspid regurgitation peak velocities were recorded.
RESULTS: There was excellent agreement between the hand held device and standard echocardiography for the evaluation of diastolic E and A waves, E/A ratio, and deceleration time with pulsed wave Doppler (intraclass correlation coefficients of 0.97, 0.93, 0.90, and 0.78, respectively). In addition, good agreement was found between continuous wave Doppler measurements of aortic ejection and tricuspid regurgitation velocities (intraclass correlation coefficients of 0.96 and 0.80). However, there was a significant difference between patients with tricuspid regurgitation measured with the hand held device (25.5%) and by standard echocardiography (65.3%), resulting in misdiagnosis of eight patients with pronounced pulmonary hypertension.
CONCLUSIONS: New hand held devices with Doppler capabilities overcome previous limitations in evaluating haemodynamic variables. With colour Doppler they are now suitable for the complete evaluation of valvar disease and diastolic function. However, important limitations remain in the evaluation of pulmonary pressures.

Entities:  

Mesh:

Year:  2003        PMID: 12923013      PMCID: PMC1767834          DOI: 10.1136/heart.89.9.1014

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  26 in total

1.  Hand-held echocardiography: revolution or hassle?

Authors:  N B Schiller
Journal:  J Am Coll Cardiol       Date:  2001-06-15       Impact factor: 24.094

2.  Recommendations for quantification of Doppler echocardiography: a report from the Doppler Quantification Task Force of the Nomenclature and Standards Committee of the American Society of Echocardiography.

Authors:  Miguel A Quiñones; Catherine M Otto; Marcus Stoddard; Alan Waggoner; William A Zoghbi
Journal:  J Am Soc Echocardiogr       Date:  2002-02       Impact factor: 5.251

3.  Spatio-temporal mapping of intracardiac pressure gradients. A solution to Euler's equation from digital postprocessing of color Doppler M-mode echocardiograms.

Authors:  J Bermejo; J C Antoranz; R Yotti; M Moreno; M A García-Fernández
Journal:  Ultrasound Med Biol       Date:  2001-05       Impact factor: 2.998

4.  Hand-carried cardiac ultrasound (HCU) device: recommendations regarding new technology. A report from the Echocardiography Task Force on New Technology of the Nomenclature and Standards Committee of the American Society of Echocardiography.

Authors:  James B Seward; Pamela S Douglas; Raimund Erbel; Richard E Kerber; Itzhak Kronzon; Harry Rakowski; L David J Sahn; Eric J Sisk; A Jamil Tajik; Samuel Wann
Journal:  J Am Soc Echocardiogr       Date:  2002-04       Impact factor: 5.251

5.  Rapid assessment of cardiac anatomy and function with a new hand-carried ultrasound device (OptiGo): a comparison with standard echocardiography.

Authors:  M Rugolotto; B S Hu; D H Liang; I Schnittger
Journal:  Eur J Echocardiogr       Date:  2001-12

6.  Physician-performed point-of-care echocardiography using a laptop platform compared with physical examination in the cardiovascular patient.

Authors:  K T Spencer; A S Anderson; A Bhargava; A C Bales; M Sorrentino; K Furlong; R M Lang
Journal:  J Am Coll Cardiol       Date:  2001-06-15       Impact factor: 24.094

Review 7.  Heart failure with preserved systolic function. A different natural history?

Authors:  M Senni; M M Redfield
Journal:  J Am Coll Cardiol       Date:  2001-11-01       Impact factor: 24.094

8.  Experience with an ultrasound stethoscope.

Authors:  Eleni C Vourvouri; Don Poldermans; Johan De Sutter; Fabiola B Sozzi; Paolo Izzo; Jos R T C Roelandt
Journal:  J Am Soc Echocardiogr       Date:  2002-01       Impact factor: 5.251

9.  Abdominal aortic aneurysm screening using a hand-held ultrasound device. "A pilot study".

Authors:  E C Vourvouri; D Poldermans; A F Schinkel; F B Sozzi; J J Bax; H van Urk; J R Roelandt
Journal:  Eur J Vasc Endovasc Surg       Date:  2001-10       Impact factor: 7.069

10.  Personal ultrasound imager: abdominal aortic aneurysm screening.

Authors:  C J Bruce; P C Spittell; S C Montgomery; K R Bailey; A J Tajik; J B Seward
Journal:  J Am Soc Echocardiogr       Date:  2000-07       Impact factor: 5.251

View more
  4 in total

1.  Ultrasound stethoscopy: a renaissance of the physical examination?

Authors:  J R T C Roelandt
Journal:  Heart       Date:  2003-09       Impact factor: 5.994

2.  Efficacy and learning curve of a hand-held echocardiography device in an oncology outpatient clinic: Expanding the use of echoscopic heart examination beyond cardiology.

Authors:  Leopoldo Pérez PéREZ DE Isla; Fernando Moreno; Jose Angel Garcia Garcia Saez; Matias Clavero; Nuno Moreno; Carlos Aguado Aguado DE LA Rosa; Jose Alberto DE Agustin; Jose Juan Gomez Gomez DE Diego; Miguel Angel Cobos; Adriana Saltijeral; Carlos Macaya; Miguel Angel Garcia-Fernandez
Journal:  Mol Clin Oncol       Date:  2015-04-09

3.  Evaluation of a hand carried cardiac ultrasound device in an outpatient cardiology clinic.

Authors:  E C Vourvouri; D Poldermans; J W Deckers; G E Parharidis; J R T C Roelandt
Journal:  Heart       Date:  2005-02       Impact factor: 5.994

4.  Hand-held echocardiography: added value in clinical cardiological assessment.

Authors:  Giovanna Giannotti; Sergio Mondillo; Maurizio Galderisi; Riccardo Barbati; Valerio Zacà; Piercarlo Ballo; Eustachio Agricola; Francesco Guerrini
Journal:  Cardiovasc Ultrasound       Date:  2005-03-24       Impact factor: 2.062

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.