Analysis of digital images acquired using different phosphor storage plates (PSPs) subjected to varying reading times and storage conditions.

M G B Q Martins1, F Haiter Neto, E J Whaites.   

Abstract

OBJECTIVES: To assess, both objectively and subjectively, the effects of different combinations of storage conditions and varying delays in reading on digital images captured using phosphor storage plates (PSPs).
METHOD: Standardized images were obtained of an aluminium step wedge and a dry mandible in acrylic using GE 1000 and PSPs from the Digora (Soredex) and DenOptix (Gendex) digital systems. Twelve plates from each system were exposed and immediately scanned to produce the baseline gold standard. The plates were re-exposed and stored using three different storage combinations, (A) ambient temperature, (R) refrigeration and (S) low humidity, and then scanned after 6 h, 12 h, 18 h, 24 h, 48 h or 72 h. Objective analysis was carried out by pixel density measurements and the data were analysed statistically using analysis of variance. Subjective analysis was carried by three oral radiologists and the results were analysed using Mann-Whitney U-test.
RESULTS: For the DenOptix system there were no differences either objectively or subjectively in the images obtained from all the plates using different storage conditions and after varying time intervals. For Digora, objective analysis showed loss of pixel density after 6 h using all storage conditions. This loss of density was also evident subjectively after 6 h on plates stored under certain storage conditions.
CONCLUSION: Storage of DenOptix PSPs for up to 72 h in different conditions does not affect the quality of the digital images. However, there is a loss of image density within 6 h when using Digora PSPs, which may affect clinical image interpretation.

Mesh:

Year:  2003        PMID: 12917285     DOI: 10.1259/dmfr/24355220

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  6 in total

1.  The effect of delayed scanning of storage phosphor plates on occlusal caries detection.

Authors:  E Sogur; B G Baksi; A Mert
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

2.  Effect of alternative photostimulable phosphor plates erasing times on subjective digital image quality.

Authors:  D P Melo; A Dos Anjos Pontual; S M de Almeida; P S F Campos; M C Alves; G M Tosoni
Journal:  Dentomaxillofac Radiol       Date:  2010-01       Impact factor: 2.419

3.  Evaluation of the effects of processing delays and protective plastic cases on image quality of a photostimulable phosphor plate system.

Authors:  Clóvis Monteiro Bramante; Alexandre Silva Bramante; Rogério Emílio de Souza; Ivaldo Gomes Moraes; Norberti Bernardineli; Roberto Brandão Garcia
Journal:  J Appl Oral Sci       Date:  2008 Sep-Oct       Impact factor: 2.698

4.  Evaluation of the Effect of Light and Scanning Time Delay on The Image Quality of Intra Oral Photostimulable Phosphor Plates.

Authors:  Amir Eskandarloo; Arman Yousefi; Setareh Soheili; Karim Ghazikhanloo; Payam Amini; Haniyeh Mohammadpoor
Journal:  Open Dent J       Date:  2017-12-29

5.  Diagnosis of Approximal Caries after Delayed Scanning of Photostimulable Phosphor Plates.

Authors:  Zahra Ghoncheh; Yasaman Kheirandish; Hanieh Kaviani; Mohammad Javad Kharazi Fard; Kurosh Sadeghi
Journal:  J Dent (Shiraz)       Date:  2017-09

6.  Comparison of the delay in processing time and protective plastic cases in two phosphor plate systems.

Authors:  Ali Murat Aktan; Mehmet Ertuğrul Ciftçi; Faruk Akgünlü
Journal:  ScientificWorldJournal       Date:  2012-11-07
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.