Temperature and humidity of the Dräger Cato anesthetic machine circuit.

Hiroki Wada1, Hideyuki Higuchi, Shinya Arimura.   

Abstract

PURPOSE: The Dräger Cato anesthetic machine (Dräger, Lübeck, Germany) effectively humidifies and warms anesthetic gases, because it has a built-in hotplate to heat the breathing system, and expired gas passes through the CO2 absorbent three times during one breath. In the present study, we measured the temperature and absolute humidity (AH) of the anesthetic circuit in the Dräger Cato machine with and without heat moisture exchangers (HME), and compared them with those in another anesthetic machine, the Aestiva/5 (Datex-Ohmeda, Helsinki, Finland).
METHODS: Forty-eight adult patients were randomly assigned to one of eight groups according to the anesthetic machine, fresh gas flow (FGF), and the use of HME (n = 6 each): Cato 0.51 x min(-1) without HME (group 1), Cato 1.01 x min(-1) without HME (group 2), Cato 0.51 x min(-1) with HME (group 3), Cato 1.01 x min(-1) with HME (group 4), Aestiva 0.51 x min(-1) without HME (group 5), Aestiva 1.01 x min(-1) without HME (group 6), Aestiva 0.51 x min(-1) with HME (group 7), and Aestiva 1.01 x min(-1) with HME (group 8). The temperature and AH of the anesthetic gases were measured with a Moiscope (S.K.I. Net, Tokyo, Japan), which was placed between the endotracheal tube and the Y-piece of the anesthetic circuit. The HME was placed between the Moiscope and the Y-piece of the anesthetic circuit. The temperature and AH of the anesthetic gases were measured at 5, 10, and 15 min and then every 15 min up to 150 min after tracheal intubation.
RESULTS: Among the groups without HME (groups 1, 2, 5, and 6), the inspired temperatures and AH in groups 1 and 2 were significantly higher than those in groups 5 and 6 at all times during the study period (P < 0.01-0.001). The inspired temperatures and AH of the groups with HME (groups 3, 4, 7, and 8), were significantly higher than those in groups 2, 5, and 6 (P < 0.01-0.001). Among the groups with HME, the AH in group 3 was significantly higher than that in group 8 until the final study period.
CONCLUSION: The present study indicates that the Dräger Cato machine was more effective in warming and humidifying respiratory gas than the Aestiva/5, and that Aestiva/5 without HME does not reach the optimal temperature and humidity ranges, even if minimal flow anesthesia (0.51 x min(-1)) is performed.

Entities:  

Mesh:

Year:  2003        PMID: 12911203     DOI: 10.1007/s00540-003-0174-8

Source DB:  PubMed          Journal:  J Anesth        ISSN: 0913-8668            Impact factor:   2.078


  4 in total

1.  Comparison of the temperature and humidity in the anesthetic breathing circuit among different anesthetic workstations: Updated guidelines for reporting parallel group randomized trials.

Authors:  Yoon Ji Choi; Sam Hong Min; Jeong Jun Park; Jang Eun Cho; Seung Zhoo Yoon; Suk Min Yoon
Journal:  Medicine (Baltimore)       Date:  2017-06       Impact factor: 1.889

2.  The Humidity in a Low-Flow Dräger Fabius Anesthesia Workstation with or without Thermal Insulation or a Heat and Moisture Exchanger: A Prospective Randomized Clinical Trial.

Authors:  Sergius A R de Oliveira; Lorena M C Lucio; Norma S P Modolo; Yoko Hayashi; Mariana G Braz; Lídia R de Carvalho; Leandro G Braz; José Reinaldo C Braz
Journal:  PLoS One       Date:  2017-01-27       Impact factor: 3.240

3.  Effects of different fresh gas flows and different anesthetics on airway temperature and humidity in surgical patients: a prospective observational study.

Authors:  Irem Gumus Ozcan; Ozkan Onal; Aysun Ozdemirkan; Ali Saltali; Mehmet Sari
Journal:  Med Gas Res       Date:  2022 Jul-Sep

4.  Insufficient Humidification of Respiratory Gases in Patients Who Are Undergoing Therapeutic Hypothermia at a Paediatric and Adult Intensive Care Unit.

Authors:  Yukari Tanaka; Sachiko Iwata; Masahiro Kinoshita; Kennosuke Tsuda; Shoichiro Tanaka; Naoko Hara; Ryota Shindou; Eimei Harada; Ryouji Kijima; Osamu Yamaga; Hitoe Ohkuma; Kazuo Ushijima; Teruo Sakamoto; Yushiro Yamashita; Osuke Iwata
Journal:  Can Respir J       Date:  2017-04-20       Impact factor: 2.409

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.