Comparison of 2-dimensional and 3-dimensional 82Rb myocardial perfusion PET imaging.

Karin Knesaurek1, Josef Machac, Borys R Krynyckyi, Orlandino D Almeida.   

Abstract

UNLABELLED: We compared 2-dimensional (2D) and 3-dimensional (3D) (82)Rb PET imaging in 3 different experiments: in a realistic heart-thorax phantom, in a uniformity-resolution phantom, and in 14 healthy volunteers.
METHODS: A nonuniform heart-thorax phantom was filled with 111 MBq of (82)Rb injected into the left ventricular (LV) wall. In the LV wall of the cardiac phantom, 3 inserts-1, 2, and 3 cm in diameter-were placed to simulate infarcts. A standard rest cardiac PET imaging protocol in 2D and 3D modes was used. Following the same protocol, a uniformity-resolution phantom with uniformly distributed activity of 1,998 MBq and 740 MBq of (82)Rb in water was used to obtain 2D PET images and 3D PET images, respectively. All 2D volunteer studies were performed by injecting 2,220 MBq of (82)Rb intravenously. For half the volunteers, 3D studies were performed with a high dose (HD) (2,220 MBq) of (82)Rb; for the remainder of the 3D studies, a low dose (LD) (740 MBq) of (82)Rb was used. In the 2D and LD 3D studies, there was a delay of 2 min and 3 min, respectively, followed by a 6-min acquisition. In the HD 3D volunteer studies, there was a delay of 5 min followed by a 6-min acquisition. Circumferential profiles of the short-axis slices and the contrast of the inserts were used to evaluate the cardiac phantom PET images. The transaxial slices from the uniformity-resolution phantom were evaluated by visual inspection and by measuring uniformity. The human studies were evaluated by measuring the contrast between LV wall and LV cavity, using linear profiles and visual analysis.
RESULTS: In the cardiac phantom study, circumferential profiles for the 2D and 3D images were similar. The contrast values for the 1-, 2-, and 3-cm inserts in the 2D study were 0.19 +/- 0.03, 0.34 +/- 0.05, and 0.61 +/- 0.03, respectively. The respective contrast values in the 3D study were 0.15 +/- 0.02, 0.36 +/- 0.04, and 0.52 +/- 0.05. In the uniformity-resolution phantom study, the coefficients of variation, calculated for a representative uniform slice, were 5.3% and 7.6% for the 2D and 3D studies, respectively. For the 7 volunteers on whom HD 3D was used, the mean 2D contrast was 0.33 +/- 0.08 and the mean HD 3D contrast was 0.35 +/- 0.08 (P = not statistically significant). For the other 7 volunteers, on whom LD 3D was used, the mean 2D contrast was 0.39 +/- 0.06 and the mean LD 3D contrast was 0.39 +/- 0.10 (P = not statistically significant). In the tomographic slices, the 2D and 3D images and polar plots were similar.
CONCLUSION: When obtained with a PET system having a high counting-rate performance, 2D and 3D (82)Rb PET cardiac images are comparable. LD 3D imaging can make (82)Rb PET cardiac imaging more affordable.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12902427

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  23 in total

1.  The new-generation positron emission tomography/computed tomography scanners: implications for cardiac imaging.

Authors:  Stephen L Bacharach
Journal:  J Nucl Cardiol       Date:  2004 Jul-Aug       Impact factor: 5.952

2.  Issues in quantification of cardiac PET studies.

Authors:  Hugo W A M de Jong; Mark Lubberink
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-03       Impact factor: 9.236

Review 3.  Present and future of clinical cardiovascular PET imaging in Europe--a position statement by the European Council of Nuclear Cardiology (ECNC).

Authors:  D Le Guludec; R Lautamäki; J Knuuti; J J Bax; F M Bengel
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-06-26       Impact factor: 9.236

Review 4.  Integration of coronary anatomy and myocardial perfusion imaging.

Authors:  Ron Blankstein; Marcelo F Di Carli
Journal:  Nat Rev Cardiol       Date:  2010-03-09       Impact factor: 32.419

5.  Status of cardiovascular PET radiation exposure and strategies for reduction: An Information Statement from the Cardiovascular PET Task Force.

Authors:  James A Case; Robert A deKemp; Piotr J Slomka; Mark F Smith; Gary V Heller; Manuel D Cerqueira
Journal:  J Nucl Cardiol       Date:  2017-05-16       Impact factor: 5.952

6.  Low-dose 3D (82)Rb PET.

Authors:  Piotr J Slomka; Daniel S Berman; Guido Germano
Journal:  J Nucl Cardiol       Date:  2012-12       Impact factor: 5.952

7.  Direct comparison between 2-dimensional and 3-dimensional PET acquisition modes for myocardial blood flow absolute quantification with O-15 water and N-13 ammonia.

Authors:  Véronique Roelants; Anne Bol; Xavier Bernard; Ann Coppens; Jacques Melin; Bernhard Gerber; Jean-Louis Vanoverschelde
Journal:  J Nucl Cardiol       Date:  2006 Mar-Apr       Impact factor: 5.952

Review 8.  PET/CT Myocardial Perfusion Imaging Acquisition and Processing: Ten Tips and Tricks to Help You Succeed.

Authors:  Talal Alnabelsi; Akanksha Thakkar; Ahmed Ibrahim Ahmed; Yushui Han; Mouaz H Al-Mallah
Journal:  Curr Cardiol Rep       Date:  2021-03-11       Impact factor: 2.931

9.  Prompt-gamma compensation in Rb-82 myocardial perfusion 3D PET/CT.

Authors:  Fabio P Esteves; Jonathan A Nye; Akbar Khan; Russell D Folks; Raghuveer K Halkar; Ernest V Garcia; David M Schuster; Stamatios Lerakis; Paolo Raggi; John R Votaw
Journal:  J Nucl Cardiol       Date:  2009-11-24       Impact factor: 5.952

10.  Repeatability of regional myocardial blood flow calculation in 82Rb PET imaging.

Authors:  Karin Knešaurek; Josef Machac; Zhuangyu Zhang
Journal:  BMC Med Phys       Date:  2009-01-29
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.