Defibrillation efficacy and pain perception of two biphasic waveforms for internal cardioversion of atrial fibrillation.

Jens Jung1, Stephen J Hahn, Armin Heisel, Axel Buob, Bernd D Schubert, Stephanos Siaplaouras.   

Abstract

UNLABELLED: Efficacy and Pain Perception of Two Biphasic Waveforms.
INTRODUCTION: We evaluated the influence of the peak voltage of waveforms used for internal cardioversion of atrial fibrillation on defibrillation efficacy and pain perception. A low peak voltage biphasic waveform generated by a 500-microF capacitor with 40% tilt was compared to a standard biphasic waveform generated by a 60-microF capacitor with 80% tilt. METHODS AND
RESULTS: In 19 patients with paroxysmal atrial fibrillation (79% male, age 55 +/- 11 years, 21% with heart disease), the atrial defibrillation threshold (ADFT) was determined during deep sedation with midazolam for both waveforms in a randomized fashion using a step-up protocol. Internal cardioversion with a single lead (shock vector: coronary sinus to right atrium) was successful in 18 (95%) of 19 patients. ADFT energy and peak voltage were significantly lower for the low-voltage waveform (2.1 +/- 2.4 J vs 3.5 +/- 3.9 J, P < 0.01; 100 +/- 53 V vs 290 +/- 149 V, P < 0.01). Sedation then was reversed with flumazenil after ADFT testing. Two shocks at the ADFT (or a 3-J shock if ADFT >3 J) were administered to the patient using each waveform in random order. Pain perception was assessed using both a visual scale and a numerical score. ADFTs were above the pain threshold in 17 (94%) of 18 patients, even though the ADFT with the 500-microF waveform was <100 V in 63% of the patients. Pain perception was comparable for both waveforms (numerical score: 6.5 +/- 2.4 vs 6.3 +/- 2.6; visual scale: 5.4 +/- 2.6 vs 5.2 +/- 3.1; P = NS, 500-microF vs 60-microF). The second shock was perceived as more painful in 88% of the patients, independent of the waveform used.
CONCLUSION: Despite a 66% lower peak voltage and a 40% lower energy, the 40% tilt, 500-microF capacitor biphasic waveform did not change the pain perceived by the patient during delivery of internal cardioversion shocks. Pain perception for internal cardioversion probably is not influenced by peak voltage alone and increases with the number of applied shocks.

Entities:  

Mesh:

Year:  2003        PMID: 12890046     DOI: 10.1046/j.1540-8167.2003.02557.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  6 in total

Review 1.  "Beauty is a light in the heart": the transformative potential of optogenetics for clinical applications in cardiovascular medicine.

Authors:  Patrick M Boyle; Thomas V Karathanos; Natalia A Trayanova
Journal:  Trends Cardiovasc Med       Date:  2014-10-16       Impact factor: 6.677

Review 2.  Invasive Management of Atrial Fibrillation and the Elderly.

Authors:  Sandeep M Patel; Samuel J Asirvatham
Journal:  J Atr Fibrillation       Date:  2010-08-23

3.  Minimum energy single-shock internal atrial defibrillation in sheep.

Authors:  Amy M Goodman; Merlise A Clyde; Donald S Burdick; Salim F Idriss; Patrick D Wolf
Journal:  J Interv Card Electrophysiol       Date:  2004-04       Impact factor: 1.900

4.  Tetanizing prepulse: A novel strategy to mitigate implantable cardioverter-defibrillator shock-related pain.

Authors:  David W Hunter; Harikrishna Tandri; Henry Halperin; Leslie Tung; Ronald D Berger
Journal:  Heart Rhythm       Date:  2016-01-06       Impact factor: 6.343

Review 5.  See the light: can optogenetics restore healthy heartbeats? And, if it can, is it really worth the effort?

Authors:  Patrick M Boyle; Emilia Entcheva; Natalia A Trayanova
Journal:  Expert Rev Cardiovasc Ther       Date:  2013-12-02

Review 6.  Innovations in Clinical Cardiac Electrophysiology: Challenges and Upcoming Solutions in 2018 and Beyond.

Authors:  Vaibhav R Vaidya; Alan Sugure; Samuel J Asirvatham
Journal:  J Innov Card Rhythm Manag       Date:  2017-12-15
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.