A Taibi1, S Fabbri, P Baldelli, C di Maggio, G Gennaro, M Marziani, A Tuffanelli, M Gambaccini.
Abstract
A dual-energy technique which employs the basis decomposition method is being investigated for application to digital mammography. A three-component phantom, made up of plexiglas, polyethylene, and water, was doubly exposed with the full-field digital mammography system manufactured by General Electric. The 'low' and 'high' energy images were recorded with a Mo/Mo anode-filter combination and a Rh/Rh combination, respectively. The total dose was kept within the acceptable levels of conventional mammography. The first hybrid images obtained with the dual-energy algorithm are presented in comparison with a conventional radiograph of the phantom. Image-quality characteristics at contrast cancellation angles between plexiglas and water are discussed. Preliminary results show that a combination of a standard Mo-anode 28 kV radiograph with a Rh-anode 49 kV radiograph provides the best compromise between image-quality and dose in the hybrid image.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12884927 DOI: 10.1088/0031-9155/48/13/307
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609