Practical considerations for the measurement of free light chains in serum.

Jillian R Tate1, Devinder Gill, Ralph Cobcroft, Peter E Hickman.   

Abstract

BACKGROUND: A new immunoassay for free light chain measurements has been reported to be useful for the diagnosis and monitoring of monoclonal light chain diseases and nonsecretory myeloma. We describe experience with and some potential pitfalls of the assay.
METHODS: The assay was assessed for precision, sample type and stability, recovery, and harmonization of results between two analyzers on which the reagents are used. Free-light-chain concentrations were measured in healthy individuals (to determine biological variation), patients with monoclonal gammopathy of undetermined significance, myeloma patients after autologous stem cell transplants, and patients with renal disease.
RESULTS: Analytical imprecision (CV) was 6-11% for kappa and lambda free-light-chain measurement and 16% for the calculated kappa/lambda ratio. Biological variation was generally insignificant compared with analytical variation. Despite the same reagent source, values were not completely harmonized between assay systems and may produce discordant free-light-chain ratios. In some patients with clinically stable myeloma, or post transplantation, or with monoclonal gammopathy of undetermined significance, free-light-chain concentration and ratio were within the population reference interval despite the presence of monoclonal intact immunoglobulin in serum. In other patients with monoclonal gammopathy of undetermined significance, values were abnormal although there was no clinical evidence of progression to multiple myeloma.
CONCLUSIONS: The use of free-light-chain measurements alone cannot differentiate some groups of patients with monoclonal gammopathy from healthy individuals. As with the introduction of any new test, it is essential that more scientific data about use of this assay in different subject groups are available so that results can be interpreted with clinical certainty.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12881439     DOI: 10.1373/49.8.1252

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  7 in total

1.  Detection of kappa and lambda light chain monoclonal proteins in human serum: automated immunoassay versus immunofixation electrophoresis.

Authors:  Troy D Jaskowski; Christine M Litwin; Harry R Hill
Journal:  Clin Vaccine Immunol       Date:  2006-02

Review 2.  Acute renal failure due to cast nephropathy in nonsecretory myeloma: a case report and review of the literature.

Authors:  S Karthik Kumar; Preet M Sohal; Harbir S Kohli; Kamal Sud; Krishan L Gupta; Kusum Joshi; Vinay Sakhuja; Vivekanand Jha
Journal:  Int Urol Nephrol       Date:  2005       Impact factor: 2.370

3.  Quantitative serum free light chain assay--analytical issues.

Authors:  Jill Tate; Sheree Bazeley; Stephen Sykes; Peter Mollee
Journal:  Clin Biochem Rev       Date:  2009-08

4.  The analytical performance evaluation of Freelite™ Human Kappa Free and Human Lambda Free on the SPAPLUS™ immunoturbidimetric analyzer.

Authors:  Kyong-Ho Cha; Yang Bo Sim; Hyojin Chae; Hae-il Park; Myungshin Kim; Yonggoo Kim
Journal:  J Clin Lab Anal       Date:  2014-01-29       Impact factor: 2.352

5.  Light chain type predicts organ involvement and survival in AL amyloidosis patients receiving stem cell transplantation.

Authors:  M Hasib Sidiqi; Mohammed A Aljama; Eli Muchtar; Francis K Buadi; Rahma Warsame; Martha Q Lacy; Angela Dispenzieri; David Dingli; Nelson Leung; Wilson I Gonsalves; Shaji K Kumar; Prashant Kapoor; Taxiarchis V Kourelis; William J Hogan; Morie A Gertz
Journal:  Blood Adv       Date:  2018-04-10

6.  The value of serum immunoglobulin free light chain assessment in patients with monoclonal gammopathies and acute renal failure.

Authors:  Mustafa Cirit; Atilla Uzüm; Pınar Ozen; Banu A Sentürk; Giray Bozkaya; Bahriye Payzin; Orçun Ural
Journal:  Turk J Haematol       Date:  2012-12-05       Impact factor: 1.831

7.  Effect of specimen type on free immunoglobulin light chains analysis on the Roche Diagnostics cobas 8000 analyzer.

Authors:  Louis S Nelson; Bryan Steussy; Cory S Morris; Matthew D Krasowski
Journal:  Springerplus       Date:  2015-12-08
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.