Deterioration in the accuracy of the pulsair non-contact tonometer with use: need for regular calibration.

P L Atkinson1, P K Wishart, J N James, S A Vernon, F Reid.   

Abstract

The Pulsair non-contact tonometer (Keeler Pulsair: Keeler UK) has been shown to be a versatile instrument particularly suitable for screening for raised intraocular pressure. Although demonstrated to be accurate initially when compared to the Goldmann applanation tonometer no study has examined its long-term accuracy. Comparisons of three Pulsair tonometers with different degrees of usage with the Goldmann tonometer are described. Measurements were obtained from 64, 116, and 223 eyes in three separate comparative studies. Correlation coefficients of between 0.78 and 0.90 were obtained, the least used instrument being significantly more accurate than the two more extensively used instruments. Taking the Goldmann tonometer as the standard tonometer, and the aim to detect intraocular pressures of greater than 21 mmHg, sensitivities of 40%, 48%, and 85% for the three Pulsair tonometers respectively were shown. The Pulsair tonometer appears liable to a long-term drift in accuracy with use, and we suggest that provision is made for the regular re-calibration of the instrument.

Mesh:

Year:  1992        PMID: 1286721     DOI: 10.1038/eye.1992.112

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  5 in total

Review 1.  Screening for glaucoma. Why is the disease underdetected?

Authors:  M W Tuck; R P Crick
Journal:  Drugs Aging       Date:  1997-01       Impact factor: 3.923

2.  Potential effects of systematic errors in intraocular pressure measurements on screening for ocular hypertension.

Authors:  M J Turner; S L Graham; A P Avolio; P Mitchell
Journal:  Eye (Lond)       Date:  2013-02-22       Impact factor: 3.775

3.  Clinical comparison of Pulsair non-contact tonometer and Goldmann applanation tonometer in Indian population.

Authors:  Shalini Mohan; Satyaprakash Tiwari; Arvind Jain; Jaya Gupta; Surendra Kumar Sachan
Journal:  J Optom       Date:  2013-07-26

4.  Glaucoma screening clinic in general practice: prevalence of occult disease, and resource implications.

Authors:  J H Sheldrick; A J Sharp
Journal:  Br J Gen Pract       Date:  1994-12       Impact factor: 5.386

5.  Design of low cost glaucoma screening.

Authors:  A G Niessen; C T Langerhorst; H C Geijssen; E L Greve
Journal:  Doc Ophthalmol       Date:  1997       Impact factor: 2.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.