Literature DB >> 12862269

Color Doppler sonographically guided transthoracic needle aspiration of lung and mediastinal masses.

Metin Gorguner1, Filiz Misirlioglu, Pinar Polat, Hasan Kaynar, Leyla Saglam, Arzu Mirici, Selami Suma.   

Abstract

OBJECTIVE: This study investigated the diagnostic value of color Doppler sonographically guided transthoracic needle aspiration in lung and mediastinal masses.
METHODS: B-mode and colorDoppler sonographic images were obtained in 48 patients with mediastinal or peripheral pulmonary tumors. Color Doppler sonography was used to show the vascular structures before the transthoracic needle aspiration procedure. It was also used to locate the needle tip during the procedure by showing the twinkling sign. This maneuver was performed with motion of the inner stylet. Pathologic and microbiological examination of the aspirates was made.
RESULTS: Vascular structures were detected in 37 cases on color Doppler images and in 10 cases on B-mode images. Similarly, the needle tip was observed in 39 cases on color Doppler images but in only 9 cases on B-mode images. No complications were observed except partial pneumothorax in 2 cases. The method had sensitivity of 90.0%, specificity of 87.5%, a positive predictive value of 97.2%, a negative predictive value of 63.6%, and diagnostic accuracy of 89.6%.
CONCLUSIONS: Color Doppler sonographically guided transthoracic needle aspiration is a safe diagnostic method in malignant lung tumors, especially peripheral tumors, because of its ability to differentiate vascular structures within a tumor before the transthoracic needle aspiration procedure. It provides additional information about the location of the needle tip.

Entities:  

Mesh:

Year:  2003        PMID: 12862269     DOI: 10.7863/jum.2003.22.7.703

Source DB:  PubMed          Journal:  J Ultrasound Med        ISSN: 0278-4297            Impact factor:   2.153


  5 in total

1.  Vibrating interventional device detection using real-time 3-D color Doppler.

Authors:  Matthew P Fronheiser; Salim F Idriss; Patrick D Wolf; Stephen W Smith
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2008       Impact factor: 2.725

2.  Diagnostic value and safety of color doppler ultrasound-guided transthoracic core needle biopsy of thoracic disease.

Authors:  Weina Huang; Libin Chen; Ning Xu; Linfeng Wang; Fang Liu; Shiyi He; Tianwen Lai; Xueqin Chen
Journal:  Biosci Rep       Date:  2019-06-07       Impact factor: 3.840

3.  Diagnostic value and complication rate of ultrasound-guided transthoracic core needle biopsy in mediastinal lesions.

Authors:  Rosen Petkov; Tzvetan Minchev; Yordanka Yamakova; Evgeni Mekov; Georgi Yankov; Danail Petrov
Journal:  PLoS One       Date:  2020-04-16       Impact factor: 3.240

4.  Ultrasonography for clinical decision-making and intervention in airway management: from the mouth to the lungs and pleurae.

Authors:  Michael S Kristensen; Wendy H Teoh; Ole Graumann; Christian B Laursen
Journal:  Insights Imaging       Date:  2014-02-12

5.  An In vitro Study of Guidewire-Related Color Doppler Twinkling Artifacts in Echocardiography.

Authors:  Shu-Ping Wu; Yi-Jui Huang; Teng-Fu Tsao; Yeu-Sheng Tyan; Chun-Hung Su
Journal:  J Med Ultrasound       Date:  2018-07-05
  5 in total