Coronary artery calcium quantification at multi-detector row CT: influence of heart rate and measurement methods on interacquisition variability initial experience.

Cheng Hong1, Kyongtae T Bae, Thomas K Pilgram, Fang Zhu.   

Abstract

PURPOSE: To assess the effect of heart rate on interacquisition variability in different coronary calcium quantification methods at multi-detector row computed tomography (CT).
MATERIALS AND METHODS: Fifty consecutive adults (39 men and 11 women; mean age, 57 years +/- 13 [SD]) with various heart rates were examined with two prospectively electrocardiographically triggered multi-detector row CT acquisitions in succession for detection and quantification of coronary artery calcification. Calcium score, volume, and mass were measured for each acquisition. Interacquisition variability was evaluated in association with heart rate and quantification method in subjects and individual coronary vessels by using t tests and analysis of variance.
RESULTS: In 37 subjects with detected calcium, interacquisition variability in mass measurement (10.4%) was significantly lower than that in score (23.9%) and volume (15.7%) measurements (P <.02). The interacquisition variability in all quantification methods was well correlated with heart rate and was considerably greater when heart rates were higher than 70 beats per minute (bpm) than when heart rates were 70 bpm or lower (P <.002). There was a clear tendency for interacquisition variability to vary by vessel (P <.01). The correlation of interacquisition variability with heart rate and a significant difference in interacquisition variability between the group with heart rates of 70 bpm or lower and the group with rates higher than 70 bpm (P <.02) were found for the left main and left anterior descending arteries but not for the circumflex and right coronary arteries.
CONCLUSION: Interacquisition variability in coronary calcium measurements at multi-detector row CT is significantly less at lower heart rates. The coronary calcium mass measurement is more reproducible than are score and volume measurements.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12832574     DOI: 10.1148/radiol.2281020685

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  8 in total

1.  Influence of iterative reconstruction on coronary calcium scores at multiple heart rates: a multivendor phantom study on state-of-the-art CT systems.

Authors:  N R van der Werf; M J Willemink; T P Willems; M J W Greuter; T Leiner
Journal:  Int J Cardiovasc Imaging       Date:  2017-12-28       Impact factor: 2.357

2.  Influence of heart rate on coronary calcium scores: a multi-manufacturer phantom study.

Authors:  N R van der Werf; M J Willemink; T P Willems; R Vliegenthart; M J W Greuter; T Leiner
Journal:  Int J Cardiovasc Imaging       Date:  2017-12-28       Impact factor: 2.357

3.  The impact of motion artifacts on the reproducibility of repeated coronary artery calcium measurements.

Authors:  Jun Horiguchi; Hiroshi Fukuda; Hideya Yamamoto; Nobuhiko Hirai; Farzana Alam; Hideaki Kakizawa; Masashi Hieda; Toshihiro Tachikake; Kazushi Marukawa; Katsuhide Ito
Journal:  Eur Radiol       Date:  2006-05-23       Impact factor: 5.315

4.  Progression of coronary artery calcification in renal transplantation and the role of secondary hyperparathyroidism and inflammation.

Authors:  Sandro Mazzaferro; Marzia Pasquali; Franco Taggi; Matteo Baldinelli; Carmina Conte; Maria Luisa Muci; Nicola Pirozzi; Iacopo Carbone; Marco Francone; Francesco Pugliese
Journal:  Clin J Am Soc Nephrol       Date:  2009-02-11       Impact factor: 8.237

5.  Evaluation of the influence of acquisition and reconstruction parameters for 16-row multidetector CT on coronary calcium scoring using a stationary and dynamic cardiac phantom.

Authors:  Philipp G C Begemann; Udo van Stevendaal; Ralph Koester; Andreas H Mahnken; Andreas Koops; Gerhard Adam; Michael Grass; Claus Nolte-Ernsting
Journal:  Eur Radiol       Date:  2007-02-01       Impact factor: 7.034

6.  Optimization of CT calcium scoring doses on the General Electric Discovery single-photon emission computed tomography/CT D670, 8-slice scanner.

Authors:  Clare Jacobs
Journal:  BJR Case Rep       Date:  2016-11-02

7.  Coronary calcium mass scores measured by identical 64-slice MDCT scanners are comparable: a cardiac phantom study.

Authors:  Hildebrand Dijkstra; Marcel J W Greuter; Jaap M Groen; Rozemarijn Vliegenthart-Proença; Klaasjan W K Renema; Frank de Lange; Matthijs Oudkerk
Journal:  Int J Cardiovasc Imaging       Date:  2009-09-19       Impact factor: 2.357

Review 8.  Coronary artery calcium score: has anything changed?

Authors:  R Marano; L Bonomo
Journal:  Radiol Med       Date:  2007-10-19       Impact factor: 6.313

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.