Literature DB >> 12826543

The EmulSiv filter removes microbial contamination from propofol but is not a substitute for aseptic technique.

Wendy C E Hall1, Donald T Jolly, Jiri Hrazdil, John C Galbraith, Maria Greacen, Alexander S Clanachan.   

Abstract

PURPOSE: To evaluate the ability of the EmulSiv filter (EF) to remove extrinsic microbial contaminants from propofol.
METHODS: Aliquots of Staphylococcus aureus (S. aureus), Candida albicans (C. albicans), Klebsiella pneumoniae (K. pneumoniae), Moraxella osloensis (M. osloensis), Enterobacter agglomerans (E. agglomerans), Escherichia coli (E. coli), Serratia marcescens (S. marcescens), Moraxella catarrhalis (M. catarrhalis), Haemophilus influenzae (H. influenzae) and Campylobacter jejuni (C. jejuni) were inoculated into vials containing 20 mL of sterile propofol. The unfiltered inoculated propofol solutions served as controls. Ten millilitres and 20 mL samples of the inoculated propofol were filtered through the EF. All solutions were then subplated onto three culture plates using a precision 1 micro L calibrated platinum loop and incubated. The number of colony forming units (CFU) were counted. Data were analyzed using a one-sample t test, and a P value of less than 0.05 was selected as the level of statistical significance.
RESULTS: The EF was able to completely remove CFU of S. aureus, C. albicans, K. pneumoniae, M. osloensis, E. agglomerans, E. coli, S. marcescens, and M. catarrhalis (P < 0.05). A small number of H. influenzae CFU were able to evade filtration in both the 10 mL and 20 mL samples. C. jejuni CFU were able to evade filtration in only the 10 mL sample.
CONCLUSIONS: The EF removes the majority of microbial contaminates from propofol with the exception of H. influenzae and C. jejuni. Although the EF is capable of removing most of the microbial contamination produced by H. influenzae and C. jejuni, a few CFU are capable of evading filtration. Consequently, even the use of a filter capable of removing microbial contaminants is not a substitute for meticulous aseptic technique and prompt administration when propofol is used.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12826543     DOI: 10.1007/BF03018637

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  2 in total

1.  Particulate contamination in intravenous drugs: coring from syringe plunger.

Authors:  S K Singhal
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2010-10

2.  Infectious Disease Risk Associated with Contaminated Propofol Anesthesia, 1989-2014(1).

Authors:  Andrés Zorrilla-Vaca; Jimmy J Arevalo; Kevin Escandón-Vargas; Daniel Soltanifar; Marek A Mirski
Journal:  Emerg Infect Dis       Date:  2016-06       Impact factor: 6.883