The effect of atrial pacing therapies on atrial tachyarrhythmia burden and frequency: results of a randomized trial in patients with bradycardia and atrial tachyarrhythmias.

Michael A Lee1, Richard Weachter, Scott Pollak, Mark S Kremers, Ajay M Naik, Russell Silverman, Joann Tuzi, Wayne Wang, Linda J Johnson, David E Euler.   

Abstract

OBJECTIVES: The Atrial Therapy Efficacy and Safety Trial (ATTEST) was a prospective, randomized study to evaluate preventive pacing and antitachycardia pacing (ATP) in patients with symptomatic atrial fibrillation (AF) or atrial tachycardia (AT).
BACKGROUND: The effect of the combination of atrial prevention and termination algorithms on AT/AF burden and frequency in pacemaker patients is unknown.
METHODS: A DDDRP pacemaker (AT500, Medtronic Inc., Minneapolis, Minnesota) with three atrial preventive pacing algorithms and two ATP algorithms was implanted in 368 patients. Patients were randomized one-month post-implant to all prevention and ATP therapies ON or OFF and followed for three months. The OFF group had DDDR pacing at a lower programmed rate of 60 ppm. The AT/AF burden and frequency were determined from daily device counters in 324 patients treated according to protocol.
RESULTS: In 17,018 episodes with stored electrograms, appropriate detection was confirmed in 17,004 (99.9%). The median percentage of atrial pacing was 98% in the ON group versus 75% in the OFF group (p < 0.001). Using device-defined criteria for successful termination, ATP terminated 8,590 (54%) of 15,789 treated episodes. The median AT/AF burden during the three-month study period was 4.2 h/month ON versus 1.1 h/month OFF (p = 0.20). The median AT/AF frequency was 1.3 episodes/month ON versus 1.2 episodes/month OFF (p = 0.65). System-related, complication-free survival at four months was 90.2% (Kaplan-Meier estimate).
CONCLUSIONS: This DDDRP pacemaker is safe, has accurate AT/AF detection, and provides ATP with 54% efficacy as defined by the device. The atrial prevention and termination therapies combined did not reduce AT/AF burden or frequency in this patient population.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12798559     DOI: 10.1016/s0735-1097(03)00426-1

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  31 in total

Review 1.  Implication from randomized trials of rate and rhythm controls on management of patients with persistent atrial fibrillation.

Authors:  Vincent E Hagens; Dirk J Van Veldhuisen; Harry J G M Crijns; Isabelle C van Gelder
Journal:  Ann Noninvasive Electrocardiol       Date:  2006-04       Impact factor: 1.468

2.  Atrial burst pacing with biphasic and monophasic waveforms for atrial fibrillation.

Authors:  Alon Barsheshet; Menachem Wakslak; Morton M Mower; Ilan Goldenberg; Burr Hall
Journal:  Ann Noninvasive Electrocardiol       Date:  2012-01       Impact factor: 1.468

Review 3.  Management of atrial fibrillation--what are the possibilities of early detection with home monitoring?

Authors:  R P Ricci; M Russo; M Santini
Journal:  Clin Res Cardiol       Date:  2006       Impact factor: 5.460

Review 4.  Atrial fibrillation: profit from cardiac pacing?

Authors:  A Yang; B Lüderitz; T Lewalter
Journal:  Z Kardiol       Date:  2005-03

Review 5.  Atrial fibrillation and conduction system disease: the roles of catheter ablation and permanent pacing.

Authors:  Anand Thiyagarajah; Dennis H Lau; Prashanthan Sanders
Journal:  J Interv Card Electrophysiol       Date:  2018-08-03       Impact factor: 1.900

Review 6.  Controversies in pacing: indications and programming.

Authors:  Anne M Gillis; Rik Willems
Journal:  Curr Cardiol Rep       Date:  2005-09       Impact factor: 2.931

7.  Intermittent vs. Continuous Anticoagulation theRapy in patiEnts with Atrial Fibrillation (iCARE-AF): a randomized pilot study.

Authors:  Stavros Stavrakis; Julie A Stoner; Joel Kardokus; Paul J Garabelli; Sunny S Po; Ralph Lazzara
Journal:  J Interv Card Electrophysiol       Date:  2016-10-01       Impact factor: 1.900

8.  The usefulness of minimal ventricular pacing and preventive AF algorithms in the treatment of PAF: the 'MinVPace' study.

Authors:  Rick A Veasey; Anita Arya; Nick Freemantle; John Silberbauer; Nikhil R Patel; Guy W Lloyd; A Neil Sulke
Journal:  J Interv Card Electrophysiol       Date:  2010-01-16       Impact factor: 1.900

9.  Effect of atrial tachyarrhythmia duration on percentage of time with atrial pacing in pacemaker patients with paroxysmal atrial fibrillation.

Authors:  Andreas Schuchert; Gaby Frost; Thomas Meinertz
Journal:  J Interv Card Electrophysiol       Date:  2004-10       Impact factor: 1.900

10.  Risk factors for bradycardia requiring pacemaker implantation in patients with atrial fibrillation.

Authors:  Tyler W Barrett; Robert L Abraham; Cathy A Jenkins; Stephan Russ; Alan B Storrow; Dawood Darbar
Journal:  Am J Cardiol       Date:  2012-07-26       Impact factor: 2.778

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.