Ultrasound axial length measurement in biphakic eyes.

Kenneth J Hoffer1.   

Abstract

To determine the calculations to correct the error in ultrasound measurement of a phakic eye with a phakic intraocular lens (IOL) (biphakic eye) to increase the accuracy of IOL power calculation should cataract surgery become necessary. Mathematical derivations were performed to calculate the error caused by a phakic IOL while measuring an eye using an average sound velocity. A simple model was developed to allow correction of the axial length (AL). If the eye is measured at an average velocity of 1555 meters/second, the following calculation can be used depending on the material of the phakic IOL: AL(corrected) = AL(1555) + X x T, where T is the central thickness of the phakic IOL and X = +0.42 (0.41 - 0.42) for poly(methyl methacrylate), -0.59 (0.56 - 0.59) for silicone, +0.11 (0.10 - 0.12) for Collamer, and +0.23 (0.23 - 0.24) for acrylic. By obtaining the dioptric power of the phakic IOL and central thickness from the manufacturer and knowing the material of the lens, a mathematical correction can be made to arrive at the true ultrasound AL in a biphakic eye.

Entities:  

Mesh:

Year:  2003        PMID: 12781283     DOI: 10.1016/s0886-3350(03)00293-1

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

Review 1.  [Phakic intraocular lenses. Current status and limitations].

Authors:  H B Dick; M Tehrani
Journal:  Ophthalmologe       Date:  2004-03       Impact factor: 1.059

2.  Comparison of preoperative and postoperative ocular biometry in eyes with phakic intraocular lens implantations.

Authors:  Joo Youn Shin; Jae Bum Lee; Kyoung Yul Seo; Eung Kweon Kim; Tae-im Kim
Journal:  Yonsei Med J       Date:  2013-09       Impact factor: 2.759

3.  Successful toric intraocular lens implantation in a patient with induced cataract and astigmatism after posterior chamber toric phakic intraocular lens implantation: a case report.

Authors:  Kazutaka Kamiya; Akio Nakamura; Hideyuki Miyake; Hiroyuki Nishimoto; Kimiya Shimizu
Journal:  J Med Case Rep       Date:  2012-04-16
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.