Reliability of corneal thickness and endothelial cell density measures.

Jason J Nichols1, Gregory M Kosunick, Mark A Bullimore.   

Abstract

PURPOSE: We evaluated the reliability and agreement between the Orbscan, an ultrasonic pachymeter (Humphrey 855), and the Konan SP 9000-LC in terms of central corneal thickness. The Konan was also used to study the reliability and agreement between endothelial cell density measures.
METHODS: Twenty-five normal subjects were examined on two occasions (mean separation = 9 +/- 5 days) by a single examiner using all three instruments for central corneal pachymetry. The Konan Center Method and a manual counting method were performed by two examiners to determine endothelial cell density. Reliability and agreement were assessed by calculating the 95% limits of agreement (LoA) and intraclass correlation coefficients (ICC).
RESULTS: For corneal pachymetry test-retest reliability, the 95% limits of agreement were -20 to +17 microm for the ultrasound, -27 to +22 microm for the Konan, and -13 to +13 microm for the Orbscan. There was fair-to-good agreement between the pachymeters (intraclass correlation coefficients range = 0.85 to 0.92). For endothelial cell density test-retest reliability, the 95% limits of agreement for the Konan Center Method was -498 to +530, and -482 to +333 cells/mm2 for examiners 1 and 2, respectively. The test-retest 95% limits of agreement for the manual overlaid grid method was -355 to +355, and -535 to +670 cells/mm2 for examiners 1 and 2, respectively.
CONCLUSIONS: The reliability and agreement of the Orbscan and Konan corneal pachymeters was good, although the reliability of the Konan for estimating endothelial cell density was fair, at best.

Entities:  

Mesh:

Year:  2003        PMID: 12777031     DOI: 10.3928/1081-597X-20030501-12

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  3 in total

1.  Effects of 5% sodium chloride ophthalmic ointment on thickness and morphology of the normal canine cornea.

Authors:  Michelle Samuel; Sara M Thomasy; Allison S Calderon; Philip H Kass; Keith Collins; Christopher J Murphy
Journal:  Vet Ophthalmol       Date:  2018-06-21       Impact factor: 1.644

2.  Intraobserver reliability of contact pachymetry in children.

Authors:  Katherine K Weise; Brett Kaminski; Michele Melia; Michael X Repka; Yasmin S Bradfield; Bradley V Davitt; David A Johnson; Raymond T Kraker; Ruth E Manny; Noelle S Matta; Susan Schloff
Journal:  J AAPOS       Date:  2013-04       Impact factor: 1.220

3.  Long-term change in central corneal thickness from a glaucoma perspective.

Authors:  Nikhil S Choudhari; Ronnie George; Ramesh Ve Sathyamangalam; Prema Raju; Rashima Asokan; Lokapavani Velumuri; Lingam Vijaya
Journal:  Indian J Ophthalmol       Date:  2013-10       Impact factor: 1.848

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.