Literature DB >> 12771836

Polyethylene damage on the nonarticular surface of modular total knee prostheses.

John M Cuckler1, Jack Lemons, J R Tamarapalli, Preston Beck.   

Abstract

Modular tibial implants submitted for retrieval analysis were examined for evidence of cold flow, wear, or polyethylene failure. All retrieved components showed areas of cold flow and wear. Significant damage, defined as pitting, gouging, or delamination was observed in 77% of the retrieved implants. Cold flow deformation commonly was observed at the junction of the polyethylene and locking mechanisms or junction of the polyethylene and the edge of the tibial tray. Control implants did not show such areas of damage. Clinical factors such as age, activity status, length of implantation, limb alignment before revision, and thickness of the tibial insert did not correlate with backside wear. No association was found between polyethylene thickness, time of implantation, limb or prosthesis alignment, or activity status or weight of the patient.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12771836     DOI: 10.1097/01.blo.0000063794.32430.05

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  6 in total

1.  Knee wear measured in retrievals: a polished tray reduces insert wear.

Authors:  Daniel J Berry; John H Currier; Michael B Mayor; John P Collier
Journal:  Clin Orthop Relat Res       Date:  2012-07       Impact factor: 4.176

2.  All-polyethylene tibial components are not inferior to metal-backed tibial components in long-term follow-up of knee arthroplasties.

Authors:  Mustafa Yassin; Avraham Garti; Moshe Weissbrot; Uzi Ashkenazi; Muhammed Khatib; Dror Robinson
Journal:  Eur J Orthop Surg Traumatol       Date:  2015-05-26

3.  Polished trays reduce backside wear independent of post location in posterior-stabilized TKAs.

Authors:  Matthew P Abdel; Mark W Gesell; Christen W Hoedt; Kathleen N Meyers; Timothy M Wright; Steven B Haas
Journal:  Clin Orthop Relat Res       Date:  2014-04-15       Impact factor: 4.176

4.  Current Total Knee Designs: Does Baseplate Roughness or Locking Mechanism Design Affect Polyethylene Backside Wear?

Authors:  Zachary W Sisko; Matthew G Teeter; Brent A Lanting; James L Howard; Richard W McCalden; Douglas D Naudie; Steven J MacDonald; Edward M Vasarhelyi
Journal:  Clin Orthop Relat Res       Date:  2017-09-13       Impact factor: 4.176

5.  Polyethylene wear is influenced by manufacturing technique in modular TKA.

Authors:  Adolph V Lombardi; Bradley S Ellison; Keith R Berend
Journal:  Clin Orthop Relat Res       Date:  2008-09-13       Impact factor: 4.176

6.  Backside wear in modern total knee designs.

Authors:  Prakash Jayabalan; Bridgette D Furman; Jocelyn M Cottrell; Timothy M Wright
Journal:  HSS J       Date:  2007-02
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.