Lauren Kaufman1, Maxim Itkin, Emma E Furth, Charmaine Stewart, Scott O Trerotola.
Abstract
A patient developed severe hepatic insufficiency after creation of a transjugular intrahepatic portosystemic shunt, which was treated unsuccessfully with a constrained Wallstent ("reducing stent"). After a failed attempt at occluding the shunt with an occlusion balloon, a detachable balloon was placed at the portal end of the constrained stent. This balloon-modified reducing stent reduced flow while maintaining shunt patency, and the patient's hepatic insufficiency improved. She went on to undergo successful liver transplantation several months later.Entities:
Mesh:
Year: 2003 PMID: 12761318 DOI: 10.1097/01.rvi.0000071094.76348.a9
Source DB: PubMed Journal: J Vasc Interv Radiol ISSN: 1051-0443 Impact factor: 3.464