Literature DB >> 12735592

Graft occlusion after deployment of the Symmetry Bypass System.

Oliver T Reuthebuch1, Alexander Kadner, Mario L Lachat, Marko I Turina.   

Abstract

Recently the Symmetry Bypass System (SJM, St. Paul, MN) became available. Now the system is frequently applied for vein-graft to aorta anastomoses in off-pump coronary artery bypass operations. This report describes a complication associated with the use of the Symmetry Bypass System (SJM) in a patient undergoing a standard off-pump coronary artery bypass procedure. A novel imaging system (SPY, Novadaq, Toronto, Canada) was applied for intraoperative assessment of graft function, and this system immediately diagnosed the occlusion of the proximal mechanical anastomosis caused by a mobile atheromatous aortic plaque.

Entities:  

Mesh:

Year:  2003        PMID: 12735592     DOI: 10.1016/s0003-4975(02)04646-5

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  2 in total

1.  Early clinical results of St. Jude Medical Symmetry aortic connector.

Authors:  Masashi Toyama; Akihiko Usui; Tomonobu Abe; Masaharu Yoshikawa; Toshiaki Akita; Yuichi Ueda
Journal:  J Artif Organs       Date:  2005       Impact factor: 1.731

Review 2.  ECM-based materials in cardiovascular applications: Inherent healing potential and augmentation of native regenerative processes.

Authors:  Anna V Piterina; Aidan J Cloonan; Claire L Meaney; Laura M Davis; Anthony Callanan; Michael T Walsh; Tim M McGloughlin
Journal:  Int J Mol Sci       Date:  2009-11-20       Impact factor: 6.208

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.