Evaluation of a flat panel digital radiographic system for low-dose portable imaging of neonates.

Ehsan Samei1, Jeanne G Hill, G Donald Frey, W Michael Southgate, Eugene Mah, David Delong.   

Abstract

The purpose of this study was to evaluate the clinical utility of an investigational flat-panel digital radiography system for low-dose portable neonatal imaging. Thirty image-pairs from neonatal intensive care unit patients were acquired with a commercial Computed Radiography system (Agfa, ADC 70), and with the investigational system (Varian, Paxscan 2520) at one-quarter of the exposure. The images were evaluated for conspicuity and localization of the endings of ancillary catheters and tubes in two observer performance experiments with three pediatric radiologists and three neonatologists serving as observers. The results indicated no statistically significant difference in diagnostic quality between the images from the investigational system and from CR. Given the investigational system's superior resolution and noise characteristics, observer results suggest that the high detective quantum efficiency of flat-panel digital radiography systems can be utilized to decrease the radiation dose/exposure to neonatal patients, although post-processing of the images remains to be optimized. The rapid availability of flat-panel images in portable imaging was found to be an added advantage for timely clinical decision-making.

Entities:  

Mesh:

Year:  2003        PMID: 12722812     DOI: 10.1118/1.1556611

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Review of a simple noise simulation technique in digital radiography.

Authors:  Rie Tanaka; Katsuhiro Ichikawa; Kosuke Matsubara; Hiroki Kawashima
Journal:  Radiol Phys Technol       Date:  2012-04-25

2.  X ray imaging goes digital.

Authors:  Kwan-Hoong Ng; Madan M Rehani
Journal:  BMJ       Date:  2006-10-14

3.  Modeling the performance characteristics of computed radiography (CR) systems.

Authors:  Srinivasan Vedantham; Andrew Karellas
Journal:  IEEE Trans Med Imaging       Date:  2010-03       Impact factor: 10.048

4.  Optimizing imaging quality and radiation dose by the age-dependent setting of tube voltage in pediatric chest digital radiography.

Authors:  Hui Guo; Wen-Ya Liu; Xiao-Ye He; Xiao-Shan Zhou; Qun-Li Zeng; Bai-Yan Li
Journal:  Korean J Radiol       Date:  2012-12-28       Impact factor: 3.500

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.