Literature DB >> 12705307

A clinical study of the efficacy of gold-tite square abutment screws in cement-retained implant restorations.

Carl J Drago1.   

Abstract

PURPOSE: The purpose of this study was to record the effectiveness of Gold-Tite square abutment screws, tightened to 35 Ncm with a torque indicator, in maintaining a clinically stable implant/abutment connection.
MATERIALS AND METHODS: The study consisted of 73 patients who were treated with 110 Osseotite implants. All patients were restored with either pre-machined titanium or customized UCLA hexed abutments. All abutments were fabricated per the Gold Standard ZR abutment design of Implant Innovations. All of the abutment screws were Gold-Tite square abutment screws. They were torqued to 35 Ncm with a torque indicator at the time of abutment connection. All crown restorations were cemented to the abutments. Patients were followed for at least 1 year post-occlusal loading.
RESULTS: Four patients with 6 implants were lost between the 6- and 12-month recall appointments. Clinical assessments of implant/restoration mobility were made by the author. One abutment screw was found to be loose at the 12-month recall appointment, representing a 99% survival rate. DISCUSSION: These results add to the growing evidence that abutment screws with enhanced surfaces may provide increased screw/implant contact, higher rotational values, and calculated preload values.
CONCLUSION: The use of the Gold-Tite square abutment screws, torqued to 35 Ncm, maintained a stable implant/abutment connection that was successful in clinical practice for this minimal evaluation period.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12705307

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  2 in total

1.  Evaluation the loosening of abutment screws in fluid contamination: an in vitro study.

Authors:  Fei Sun; Wei Cheng; Bao-Hong Zhao; Gui-Qiu Song; Zeng Lin
Journal:  Sci Rep       Date:  2022-06-24       Impact factor: 4.996

2.  Abrasion of abutment screw coated with TiN.

Authors:  Seok-Won Jung; Mee-Kyoung Son; Chae-Heon Chung; Hee-Jung Kim
Journal:  J Adv Prosthodont       Date:  2009-07-31       Impact factor: 1.904