Antonello De Martino1, Yong-Ki Kim, Enric Garcia-Caurel, Blandine Laude, Bernard Drévillon.
Abstract
We demonstrate a Mueller polarimeter in which the polarization-state generator and analyzer are both composed of a linear polarizer and two liquid-crystal variable retarders. The polarimeter is designed to optimize the accuracy of the final results by minimization of the condition numbers of the modulation and analysis matrices. The polarimeter calibration, a difficult task by conventional procedures, is achieved easily by use of the eigenvalue method of Compain et al. [Appl. Opt. 38, 3490 (1999)]. The overall polarimeter performance is tested with a linear polarizer at various angles and a compensator at various retardations.Entities:
Year: 2003 PMID: 12703918 DOI: 10.1364/ol.28.000616
Source DB: PubMed Journal: Opt Lett ISSN: 0146-9592 Impact factor: 3.776