Comparison of thermodilution bolus cardiac output and Doppler cardiac output in the early post-cardiopulmonary bypass period.

Xiaoqin Zhao1, John S Mashikian, Pete Panzica, Adam Lerner, Kyung W Park, Mark E Comunale.   

Abstract

OBJECTIVE: To evaluate the accuracy of measuring cardiac output (CO) in the early post-cardiopulmonary bypass (CPB) period by comparing thermodilution with Doppler methods.
DESIGN: Prospective and blinded human trial.
SETTING: Academic medical center. PARTICIPANTS: Thirty adult patients undergoing elective coronary artery bypass graft surgery.
MEASUREMENTS AND MAIN RESULTS: Thermodilution CO (TCO) was obtained in triplicate. Doppler CO (DCO) in triplicate was obtained at the left ventricular outflow tract (LVOT), aortic valve (AV), and right ventricular outflow tract (RVOT). CO measurements were made (1). before CPB (baseline), (2). immediately after CPB, (3).15 minutes after CPB, and (4). 30 minutes after CPB. Before CPB, the DCO at LVOT, RVOT, and AV showed good correlations (r = 0.87, r = 0.88, and r = 0.84, respectively) with TCO. Bias analysis showed no significant difference among TCO and 3 DCOs (p > 0.05 each). Correlation between DCO and TCO decreased but remained significant after CPB (r between 0.57 and 0.85, p < 0.001). The bias among TCO and each of the DCOs at the LVOT, RVOT, and AV increased immediately after CPB (p < 0.01, p < 0.01, and p < 0.05, respectively) and remained significant at 15 minutes and 30 minutes post-CPB except for DCO at the AV. TCO exceeded DCO by 0.44 to 0.72 L/min immediately after CPB. The CO measured by both thermodilution and Doppler methods gradually decreased over time post-CPB. The decrease in CO was significant at 30 minutes post-CPB (p < 0.01).
CONCLUSION: This study adds further support that DCO is a clinically acceptable method to accurately assess the CO in patients even during periods of uneven regional body temperatures as may occur in the early post-CPB period. Copyright 2003 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2003        PMID: 12698401     DOI: 10.1053/jcan.2003.46

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  4 in total

1.  Can intraoperative TEE correctly measure residual shunt after surgical repair of ventricular septal defects?

Authors:  Satoshi Kurokawa; Takayuki Honma; Miki Taneoka; Hidekazu Imai; Hiroshi Baba; Minoru Nomura
Journal:  J Anesth       Date:  2010-03-13       Impact factor: 2.078

Review 2.  Systematic review of cardiac output measurements by echocardiography vs. thermodilution: the techniques are not interchangeable.

Authors:  Mik Wetterslev; Hasse Møller-Sørensen; Rasmus Rothmann Johansen; Anders Perner
Journal:  Intensive Care Med       Date:  2016-03-01       Impact factor: 17.440

3.  Reduced Right Ventricular Function Predicts Long-Term Cardiac Re-Hospitalization after Cardiac Surgery.

Authors:  Leela K Lella; Virna L Sales; Yulia Goldsmith; Jacqueline Chan; Marina Iskandir; Iosif Gulkarov; Anthony Tortolani; Sorin J Brener; Terrence J Sacchi; John F Heitner
Journal:  PLoS One       Date:  2015-07-21       Impact factor: 3.240

4.  Cardiac output measurements via echocardiography versus thermodilution: A systematic review and meta-analysis.

Authors:  Yun Zhang; Yan Wang; Jing Shi; Zhiqiang Hua; Jinyu Xu
Journal:  PLoS One       Date:  2019-10-03       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.