Literature DB >> 12677113

Tolerable hearing aid delays. III. Effects on speech production and perception of across-frequency variation in delay.

Michael A Stone1, Brian C J Moore.   

Abstract

OBJECTIVE: Signal processing strategies that attempt to mimic the frequency resolution of the healthy cochlea require finer frequency resolution at low frequencies than at high. The filtering required will cause more delay to the low-frequency end of the spectrum than to the high-frequency end, which may have disturbing effects. In a real-time application, making the delay constant across frequency would require the higher frequency components to be subjected to a compensating delay, leading to a larger overall delay, which may also give rise to disturbing effects (Stone & Moore, 1999, 2002). This experiment assessed the effects of frequency-dependent delay on subjective and objective measures of speech production and perception.
DESIGN: Digital filtering was used to introduce across-frequency delays of 0, 4, 9, 15, or 24 msec. The low frequencies were always more delayed than the high frequencies, with a smooth variation in delay between the low and the high frequencies. In addition, there was a 2.5 msec delay that did not vary with frequency arising from the processes of analog-digital-analog conversion and equalization. Ten subjects with symmetric, bilateral, moderate hearing impairment of cochlear origin were fitted binaurally with Phonak PicoNet2 BTE linear hearing aids. The aids were programmed to provide insertion gains closely approximating the NAL (RP) (Byrne & Dillon, 1986) prescription for each subject and ear. The output of an omnidirectional microphone, attached on top of each aid, fed a digital filtering unit that introduced the across-frequency delay. Audio shoes reintroduced the signal back into the BTE aids. Subjects were tested in a counterbalanced order for their identification of vowel-consonant-vowel (VCV) syllables. In a second session, the same subjects spoke from a script: their utterances were recorded on tape, speech production rates were measured and subjective ratings of the disturbance of the delay were obtained. Subjects required some training to recognize the effects of the delay and rate them consistently.
RESULTS: Subjective disturbance increased progressively with increasing across-frequency delay. A 9-msec delay was significantly more disturbing than a 0-msec delay. A delay of about 20 msec led to a mean rating of "disturbing." VCV identification decreased significantly once the across-frequency delay was 15 msec or greater. However, word production rates were not significantly affected by across-frequency delay over the range tested.
CONCLUSIONS: Relatively small across-frequency delays (9 to 15 msec) give rise to significant changes in speech identification and subjective disturbance. A delay of 9 to 15 msec, constant across frequency, would have a smaller effect (Stone & Moore, 1999). It appears desirable to compensate, at least in part, for across-frequency delays introduced by any processing.

Entities:  

Mesh:

Year:  2003        PMID: 12677113     DOI: 10.1097/01.AUD.0000058106.68049.9C

Source DB:  PubMed          Journal:  Ear Hear        ISSN: 0196-0202            Impact factor:   3.570


  14 in total

Review 1.  Challenges and recent developments in hearing aids. Part I. Speech understanding in noise, microphone technologies and noise reduction algorithms.

Authors:  King Chung
Journal:  Trends Amplif       Date:  2004

Review 2.  Principles of digital dynamic-range compression.

Authors:  James M Kates
Journal:  Trends Amplif       Date:  2005

3.  Analog-to-digital conversion to accommodate the dynamics of live music in hearing instruments.

Authors:  Neil S Hockley; Frauke Bahlmann; Bernadette Fulton
Journal:  Trends Amplif       Date:  2012-09

4.  Bimodal Cochlear Implant Listeners' Ability to Perceive Minimal Audible Angle Differences.

Authors:  Ashley Zaleski-King; Matthew J Goupell; Dragana Barac-Cikoja; Matthew Bakke
Journal:  J Am Acad Audiol       Date:  2018-11-12       Impact factor: 1.664

5.  Open Speech Platform: Democratizing Hearing Aid Research.

Authors:  Dhiman Sengupta; Arthur Boothroyd; Tamara Zubatiy; Cagri Yalcin; Dezhi Hong; Sean K Hamilton; Rajesh Gupta; Harinath Garudadri
Journal:  Int Conf Pervasive Comput Technol Healthc       Date:  2020-05

6.  The effect of hearing impairment on the identification of speech that is modulated synchronously or asynchronously across frequency.

Authors:  Joseph W Hall; Emily Buss; John H Grose
Journal:  J Acoust Soc Am       Date:  2008-02       Impact factor: 1.840

7.  The speech intelligibility benefit of a unilateral wireless system for hearing-impaired adults.

Authors:  William M Whitmer; Christopher G Brennan-Jones; Michael A Akeroyd
Journal:  Int J Audiol       Date:  2011-09-20       Impact factor: 2.117

Review 8.  Effects of bandwidth, compression speed, and gain at high frequencies on preferences for amplified music.

Authors:  Brian C J Moore
Journal:  Trends Amplif       Date:  2012-11-19

9.  Real-time contrast enhancement to improve speech recognition.

Authors:  Joshua M Alexander; Rick L Jenison; Keith R Kluender
Journal:  PLoS One       Date:  2011-09-19       Impact factor: 3.240

10.  Slow Cortical Potentials and Amplification-Part II: Acoustic Measures.

Authors:  Lorienne M Jenstad; Susan Marynewich; David R Stapells
Journal:  Int J Otolaryngol       Date:  2012-10-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.