Literature DB >> 12667005

Pilot external quality assessment survey for post-market vigilance of in vitro diagnostic medical devices and investigation of trueness of participants' results.

Linda M Thienpont1, Dietmar Stöckl, Josef Kratochvíla, Bedrich Friedecký, Marek Budina.   

Abstract

The Czech External Quality Assessment Scheme organized a survey using 14 fresh-frozen sera targeted for cholesterol and glucose by reference measurement procedures. The objective was to investigate whether it could fulfil a post-market vigilance function for in vitro diagnostic medical devices and assess trueness of participants' results. It revealed a mean bias of +5.1% for cholesterol and +3.7% for glucose (n approximately 150). However, the bias source (manufacturer or laboratory) could not be identified unequivocally because of the lack of homogeneous groups. This was due to the fact that laboratories mainly used reagents from manufacturers that do not market instruments or combined calibrators and reagents from different sources. Consequently, these habits did not allow the survey to fulfil the vigilance function. On the other hand, we were able to show the individual participants results for patient samples deviating from the true value (deviations >10% in approximately 20% of the laboratories). However, again, the survey failed in problem-solving via peer-group evaluation, even for participants that applied homogeneous tests. If other European schemes confirm this outcome, cooperation and/or participation of manufacturers may be the solution. The survey pointed out to the other participants, interchanging instrument, reagent and calibrator, that they are themselves responsible for the problems shown and hence also for problem-solving.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12667005     DOI: 10.1515/CCLM.2003.030

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  4 in total

1.  Measurements for 8 common analytes in native sera identify inadequate standardization among 6 routine laboratory assays.

Authors:  Hedwig C M Stepman; Ulla Tiikkainen; Dietmar Stöckl; Hubert W Vesper; Selvin H Edwards; Harri Laitinen; Jonna Pelanti; Linda M Thienpont
Journal:  Clin Chem       Date:  2014-03-31       Impact factor: 8.327

2.  Traceability, reference systems and result comparability.

Authors:  Mauro Panteghini
Journal:  Clin Biochem Rev       Date:  2007-08

3.  Trueness assessment for serum glucose measurement using commercial systems through the preparation of commutable reference materials.

Authors:  Changyu Xia; Ou Liu; Lanzhen Wang; Guobing Xu
Journal:  Ann Lab Med       Date:  2012-06-20       Impact factor: 3.464

Review 4.  Role and Responsibilities of Laboratory Medicine Specialists in the Verification OF Metrological Traceability of in vitro Medical Diagnostics.

Authors:  Federica Braga; Ilenia Infusino; Mauro Panteghini
Journal:  J Med Biochem       Date:  2015-07-14       Impact factor: 3.402

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.