Decrease in power output of new light-emitting diode (LED) curing devices with increasing distance to filling surface.

Gerrit R Meyer1, Claus-Peter Ernst, Brita Willershausen.   

Abstract

PURPOSE: This study compares the percentage decrease in power output of light-emitting diode (LED) and quartz-tungsten-halogen (QTH) curing devices at different distances between the light guide tip and a radiometer.
MATERIALS AND METHODS: Three LED curing devices (Elipar FreeLight/3M ESPE, Luxomax/Akeda, e-Light/GC) were compared to two QTH curing devices (Elipar TriLight/3M ESPE, Optilux 501 conventional and Turbo light guides/Kerr-Demetron). Power density was measured with a Fieldmaster GS/Coherent unit (Sensor LM-3 HTD) at increasing distances from the light guide tip (0 to 20 mm at 1-mm increments, n=6).
RESULTS: The mean decrease in power density available for curing at a distance of 10 mm between the light guide tip and the radiometer was 68% for the FreeLight, 83% for the e-Light, 42% for the Luxomax, 38% for the TriLight, 33% for the Optilux with the standard light guide, and 44% for the Optilux with the turbo light guide. The power density of the Luxomax was only 40% of that of the other LED curing devices at 0 mm distance.
CONCLUSION: The blue LED curing devices Elipar FreeLight and GC e-Light showed a significantly higher percentage decrease in power output at a distance of 10 mm from the light guide tip to the radiometer compared to the QTH curing devices. Therefore, although blue LED curing devices might have the same curing potential compared to a QTH device when placed in direct contact to a resin composite, blue LED curing devices may not provide a sufficient cure when placed at a clinically relevant distance of 10 mm to the resin composite surface.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12666755

Source DB:  PubMed          Journal:  J Adhes Dent        ISSN: 1461-5185            Impact factor:   2.359


  6 in total

1.  Influence of physical assessment of different light-curing units on irradiance and composite microhardness top/bottom ratio.

Authors:  Susana Morimoto; Renata Aló Maluza Zanini; Josete Barbosa Cruz Meira; Carlos Martins Agra; Fernanda Calabró Calheiros; Denis Yudi Nagase
Journal:  Odontology       Date:  2016-01-13       Impact factor: 2.634

2.  Effect of power density of curing unit, exposure duration, and light guide distance on composite depth of cure.

Authors:  Anders Lindberg; Anne Peutzfeldt; Jan W V van Dijken
Journal:  Clin Oral Investig       Date:  2005-04-07       Impact factor: 3.573

3.  Temperature rise and degree of photopolymerization conversion of nanocomposites and conventional dental composites.

Authors:  Mohammad Atai; Fariba Motevasselian
Journal:  Clin Oral Investig       Date:  2008-12-16       Impact factor: 3.573

4.  Depth of cure and mechanical properties of nano-hybrid resin-based composites with novel and conventional matrix formulation.

Authors:  Karina E Frauscher; Nicoleta Ilie
Journal:  Clin Oral Investig       Date:  2011-12-02       Impact factor: 3.573

5.  Degree of conversion of nano-hybrid resin-based composites with novel and conventional matrix formulation.

Authors:  Karina E Frauscher; Nicoleta Ilie
Journal:  Clin Oral Investig       Date:  2012-05-03       Impact factor: 3.573

6.  Impact of light-curing distance on the effectiveness of cure of bulk-fill resin-based composites.

Authors:  Rana Abdelbaset Diab; Adrian Ujin Yap; Maria Angela Garcia Gonzalez; Noor Azlin Yahya
Journal:  Saudi Dent J       Date:  2021-01-14
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.