Literature DB >> 12663017

Clinical assessment of the Zeiss IOLMaster.

Ingrid Kielhorn1, Madhavan S Rajan, Peter M Tesha, Visva Raja Subryan, John A Bell.   

Abstract

PURPOSE: To compare axial lengths and intraocular lens (IOL) power calculations obtained with the Zeiss IOLMaster by 2 observers with different training grades and to assess the agreement between optical and acoustic biometry.
SETTING: Department of Ophthalmology, Pembury Hospital, The Laser Eye Clinic, Kent, United Kingdom.
METHODS: In this prospective study, optical biometry was performed twice in 79 patients: first by a qualified ophthalmic nurse and second by 1 of several health care assistants without formal nursing training. In the same group of patients, the qualified nurse also obtained ultrasonic measurements to compare the 2 clinical methods.
RESULTS: There was a high correlation coefficient for axial length and IOL power measurements (0.99 each) between the experienced examiner and the inexperienced examiners. The coefficient of repeatability (2 SD) was 0.58 diopter (D) for IOL power (mean 0.04 D) and 0.07 mm for the axial length (mean 0.00 mm). The correlation coefficient of IOLMaster measurements with A-scan measurements done by the same experienced ophthalmic nurse was 0.97 for IOL power and 0.98 for axial length. The limits of agreement (mean +/- 2 SD) were between +0.50 and -2.40 D for IOL power calculations and between +0.7 mm and -0.2 mm for axial length measurements.
CONCLUSIONS: Optical biometry showed excellent repeatability using different examiners regardless of their medical training. The agreement with ultrasonic measurements and the precision of the method need further evaluation.

Entities:  

Mesh:

Year:  2003        PMID: 12663017     DOI: 10.1016/s0886-3350(02)01819-9

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  11 in total

1.  Relationship between the retinal thickness of the macula and the difference in axial length.

Authors:  Tetsuo Ueda; Yoshiaki Nawa; Yoshiaki Hara
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-08-13       Impact factor: 3.117

Review 2.  [Apodized diffractive optic. New concept in multifocal lens technology].

Authors:  T Kohnen; V Derhartunian
Journal:  Ophthalmologe       Date:  2007-10       Impact factor: 1.059

3.  How to avoid mistakes in biometry.

Authors:  Nick Astbury; Balasubramanya Ramamurthy
Journal:  Community Eye Health       Date:  2006-12

4.  Agreement analysis of LENSTAR with other techniques of biometry.

Authors:  S Jasvinder; T F Khang; K K S Sarinder; V P Loo; V Subrayan
Journal:  Eye (Lond)       Date:  2011-03-11       Impact factor: 3.775

5.  Accutome PachPen handheld ultrasonic pachymeter: intraobserver repeatability and interobserver reproducibility by personnel of different training grades.

Authors:  Mohammadreza Peyman; Lai Yong Tai; Keat Ween Khaw; Choung Min Ng; Maung Maung Win; Visvaraja Subrayan
Journal:  Int Ophthalmol       Date:  2014-09-05       Impact factor: 2.031

6.  A Model to Predict Postoperative Axial Length in Children Undergoing Bilateral Cataract Surgery With Primary Intraocular Lens Implantation.

Authors:  Rupal H Trivedi; Eliza Barnwell; Bethany Wolf; M Edward Wilson
Journal:  Am J Ophthalmol       Date:  2019-05-10       Impact factor: 5.258

7.  [Measurement of accommodation using optical biometry].

Authors:  A Nurispahic; K Kotliar; I Lanzl
Journal:  Ophthalmologe       Date:  2008-04       Impact factor: 1.059

8.  Comparison of two optical biometers in intraocular lens power calculation.

Authors:  Sheng Hui; Lu Yi
Journal:  Indian J Ophthalmol       Date:  2014-09       Impact factor: 1.848

9.  Effect of pupillary dilation on Haigis formula-calculated intraocular lens power measurement by using optical biometry.

Authors:  Bharkbhum Khambhiphant; Suganlaya Sasiwilasagorn; Nattida Chatbunchachai; Krit Pongpirul
Journal:  Clin Ophthalmol       Date:  2016-07-27

10.  Meta-analysis of optical low-coherence reflectometry versus partial coherence interferometry biometry.

Authors:  Jinhai Huang; Colm McAlinden; Yingying Huang; Daizong Wen; Giacomo Savini; Ruixue Tu; Qinmei Wang
Journal:  Sci Rep       Date:  2017-02-24       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.