Physical aspects of dynamic stereotactic radiosurgery with very small photon beams (1.5 and 3 mm in diameter).

Kamen A Paskalev1, Jan P Seuntjens, Horacio J Patrocinio, Ervin B Podgorsak.   

Abstract

Stereotactic radiosurgery is often used for treating functional disorders. For some of these disorders, the size of the target can be on the order of a millimeter and the radiation dose required for treatment on the order of 80 Gy. The very small radiation field and high prescribed dose present a difficult challenge in beam calibration, dose distribution calculation, and dose delivery. In this work the dose distribution for dynamic stereotactic radiosurgery, carried out with 1.5 and 3 mm circular fields, was studied. A 10 MV beam from a Clinac-18 linac (Varian, Palo Alto, CA) was used as the radiation source. The BEAM/EGS4 Monte Carlo code was used to model the treatment head of the machine along with the small-field collimators. The models were validated with the EGSnrc code, first through a calculation of percent depth doses (PDD) and dose profiles in a water phantom for the two small stationary circular beams and then through a comparison of the calculated with measured PDD and profile data. The three-dimensional (3-D) dose distributions for the dynamic rotation with the two small radiosurgical fields were calculated in a spherical water phantom using a modified version of the fast XVMC Monte Carlo code and the validated models of the machine. The dose distributions in a horizontal plane at the isocenter of the linac were measured with low-speed radiographic film. The maximum sizes of the Monte Carlo-calculated 50% isodose surfaces in this horizontal plane were 2.3 mm for the 1.5 mm diameter beam and 3.8 mm for the 3 mm diameter beam. The maximum discrepancies between the 50% isodose surface on the film and the 50% Monte Carlo-calculated isodose surfaces were 0.3 mm for both the 1.5 and 3 mm beams. In addition, the displacement of the delivered dose distributions with respect to the laser-defined isocenter of the machine was studied. The results showed that dynamic radiosurgery with very small beams has a potential for clinical use.

Entities:  

Mesh:

Year:  2003        PMID: 12607827     DOI: 10.1118/1.1536290

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  PRESAGE 3D dosimetry accurately measures Gamma Knife output factors.

Authors:  Slade J Klawikowski; James N Yang; John Adamovics; Geoffrey S Ibbott
Journal:  Phys Med Biol       Date:  2014-11-04       Impact factor: 3.609

2.  On the quantification of the dosimetric accuracy of collapsed cone convolution superposition (CCCS) algorithm for small lung volumes using IMRT.

Authors:  Oscar I Calvo; Alonso N Gutiérrez; Sotirios Stathakis; Carlos Esquivel; Nikos Papanikolaou
Journal:  J Appl Clin Med Phys       Date:  2012-05-10       Impact factor: 2.102

3.  A comparison of the convolution and TMR10 treatment planning algorithms for Gamma Knife® radiosurgery.

Authors:  Peter Fallows; Gavin Wright; Natalie Harrold; Peter Bownes
Journal:  J Radiosurg SBRT       Date:  2018

4.  Dosimetric algorithm to reproduce isodose curves obtained from a LINAC.

Authors:  Julio Cesar Estrada Espinosa; Segundo Agustín Martínez Ovalle; Cinthia Kotzian Pereira Benavides
Journal:  Comput Math Methods Med       Date:  2014-06-18       Impact factor: 2.238

5.  Comparison of trigeminal neuralgia radiosurgery plans using two film detectors for the commissioning of small photon beams.

Authors:  Karina P Esparza-Moreno; Olivia A García-Garduño; Paola Ballesteros-Zebadúa; José M Lárraga-Gutiérrez; Sergio Moreno-Jiménez; Miguel A Celis-Lopez
Journal:  J Appl Clin Med Phys       Date:  2013-11-04       Impact factor: 2.102

6.  Clinical Evaluation of a Two-dimensional Liquid-Filled Ion chamber Detector Array for Verification of High Modulation Small Fields in Radiotherapy.

Authors:  Miljenko Markovic; Ganesh Narayanasamy; Sotirios Stathakis; Panayiotis Mavroidis; Ines-Ana Jurkovic; Daniel Saenz; Niko Papanikolaou
Journal:  J Med Phys       Date:  2019 Apr-Jun

7.  An Experimental Slope Method for a More Accurate Measurement of Relative Radiation Doses using Radiographic and Radiochromic Films and Its Application to Megavoltage Small-Field Dosimetry.

Authors:  Raghavendra Holla; D Khanna; Bhaskaran K Pillai; K V Jafar Ali; P S Renil Mon; C O Clinto; Tharmarnadar Ganesh
Journal:  J Med Phys       Date:  2019 Jul-Sep
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.