[Theory and software of the hexapod external fixator].

K Seide1, J Wolnack, N Weinrich, C Jürgens.   

Abstract

Using hexapod kinematics (Stewart platform), an external fixator was developed that can be adjusted in all six spatial degrees of freedom by means of six linear adjusting elements. With such a system, any desired three-dimensional bone movements, for example, fracture reduction or deformity correction, can be effected exactly, without having temporarily to compromise any stability already achieved or to rearrange the construction during treatment. Since the introduction of the hexapod principle by Stewart in 1965, computerized control has been deemed necessary for its application. This paper describes the mathematical basis and the software developed for clinical use. Mathematical procedures are needed for the calculation of the inverse and forward kinematics of the hexapod, and for the description of three-dimensional movements.

Entities:  

Mesh:

Year:  2002        PMID: 12585052     DOI: 10.1515/bmte.2002.47.12.326

Source DB:  PubMed          Journal:  Biomed Tech (Berl)        ISSN: 0013-5585            Impact factor:   1.411