Literature DB >> 12578083

The calibration and validation of a phase-modulated near-infrared cerebral oximeter.

W J Levy1, J Carpenter, R M Fairman, M A Golden, E Zager.   

Abstract

OBJECTIVE: This study was undertaken to compare the cerebral oxygenation measured by an experimental phase-modulated near-infrared (NIR) spectroscopy system with capillary saturation estimated from jugular venous oxygen saturation.
METHODS: Jugular venous catheters were placed in 30 patients undergoing carotid endarterectomy and 194 measurements of venous oxygen saturation were obtained intra operatively. Simultaneous measurement of optical path length at 754, 785, and 816 nm was performed using a phase-modulated near-infrared spectroscopy system. Optical calibration was performed using both an optical bench and a scattering mold. Hemoglobin saturation was calculated from NIR measurements using equations derived from diffusion theory. Capillary saturation was calculated from the arterial and venous saturations.
RESULTS: Jugular venous saturations ranged from 41 to 92%. When calibrated using the optical bench, the NIR estimates of hemoglobin saturation deviated from estimated capillary values by an average of 2.6% bias and 4.3% deviation. No systematic bias was noted. NIR values derived from mold calibration were less accurate and precise (4.6% bias and 6.9% deviation.) Use of the initial venous sample as an in vivo calibration improved the accuracy of the mold calibration but did not alter the performance of the bench calibration.
CONCLUSIONS: Under the conditions tested, an experimental phase-modulated near-infrared spectroscopy system calibrated using an optical bench agreed with capillary saturation estimated from jugular venous samples. Further work is necessary to demonstrate valid performance of the system under other conditions.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 12578083     DOI: 10.1023/a:1009984204752

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  10 in total

1.  Quantitation of time- and frequency-resolved optical spectra for the determination of tissue oxygenation.

Authors:  E M Sevick; B Chance; J Leigh; S Nioka; M Maris
Journal:  Anal Biochem       Date:  1991-06       Impact factor: 3.365

2.  Effect of optode separation on brain penetration in adults.

Authors:  D N Harris; S M Bailey; F Cowans; D Wertheim
Journal:  Adv Exp Med Biol       Date:  1996       Impact factor: 2.622

3.  The contribution of extracranial blood oxygenation on near-infrared spectroscopy during carotid thrombendarterectomy.

Authors:  G Grubhofer; A Lassnigg; F Manlik; E Marx; W Trubel; M Hiesmayr
Journal:  Anaesthesia       Date:  1997-02       Impact factor: 6.955

4.  Bilateral jugular venous blood flow by thermal dilution.

Authors:  E M Wilson; J H Halsey
Journal:  Stroke       Date:  1970 Sep-Oct       Impact factor: 7.914

5.  The influence of demographic factors on phase-modulated spectroscopy in adults.

Authors:  W J Levy
Journal:  Philos Trans R Soc Lond B Biol Sci       Date:  1997-06-29       Impact factor: 6.237

6.  Regional cerebrovascular oxygen saturation measured by optical spectroscopy in humans.

Authors:  P W McCormick; M Stewart; M G Goetting; G Balakrishnan
Journal:  Stroke       Date:  1991-05       Impact factor: 7.914

7.  Near-infrared measurement of cerebral oxygenation. Correlation with electroencephalographic ischemia during ventricular fibrillation.

Authors:  W J Levy; S Levin; B Chance
Journal:  Anesthesiology       Date:  1995-10       Impact factor: 7.892

8.  The influence of carbon dioxide and body position on near-infrared spectroscopic assessment of cerebral hemoglobin oxygen saturation.

Authors:  V Pollard; D S Prough; A E DeMelo; D J Deyo; T Uchida; R Widman
Journal:  Anesth Analg       Date:  1996-02       Impact factor: 5.108

9.  Kinetics of cerebral deoxygenation during deep hypothermic circulatory arrest in neonates.

Authors:  C D Kurth; J M Steven; S C Nicolson; B Chance; M Delivoria-Papadopoulos
Journal:  Anesthesiology       Date:  1992-10       Impact factor: 7.892

10.  Internal and external carotid contributions to near-infrared spectroscopy during carotid endarterectomy.

Authors:  J M Lam; P Smielewski; P al-Rawi; P Griffiths; J D Pickard; P J Kirkpatrick
Journal:  Stroke       Date:  1997-05       Impact factor: 7.914

  10 in total
  2 in total

Review 1.  In vivo validation of cerebral near-infrared spectroscopy: a review.

Authors:  Amalie la Cour; Gorm Greisen; Simon Hyttel-Sorensen
Journal:  Neurophotonics       Date:  2018-11-27       Impact factor: 3.593

2.  Estimation of jugular venous O2 saturation from cerebral oximetry or arterial O2 saturation during isocapnic hypoxia.

Authors:  M B Kim; D S Ward; C R Cartwright; J Kolano; S Chlebowski; L C Henson
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.