Four-year study of cobalt and chromium blood levels in patients managed with two different metal-on-metal total hip replacements.

Christian Lhotka1, Thomas Szekeres, Ilse Steffan, Klaus Zhuber, Karl Zweymüller.   

Abstract

BACKGROUND: Metal-on-metal total hip prostheses will produce a certain amount of wear debris. This results in increased whole-blood metal levels, which may cause adverse effects. It is not known to what extent the problem has been overcome by advances in alloy technology.
METHODS: In 259 patients who with total hip replacement, blood cobalt and chromium concentrations were measured with atomic absorption spectrophotometry over a period of four years after arthroplasty. Of the patients enrolled in the study, 131 had been managed with a METASUL cobalt-chromium alloy metal-on-metal bearing combination, while 128 had been given a SIKOMET-SM21 cobalt-chromium alloy metal-on-metal combination. The control group consisted of 31 age- and gender-matched subjects.
RESULTS: Compared with the controls, all the patients had higher cobalt and chromium levels. Cobalt concentrations were up to 50 times higher, while chromium concentrations were up to 100 times higher.
CONCLUSIONS: Both systems showed evidence, in the whole-blood samples, of wear debris production by the implants. Therefore, patients managed with metal-on-metal bearing combinations should be carefully monitored in order to ensure that any local or systemic complications are detected early on.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12568948     DOI: 10.1016/S0736-0266(02)00152-3

Source DB:  PubMed          Journal:  J Orthop Res        ISSN: 0736-0266            Impact factor:   3.494


  27 in total

1.  Factors to consider in joint prosthesis systems.

Authors:  Larry M Wolford
Journal:  Proc (Bayl Univ Med Cent)       Date:  2006-07

2.  Serum ion level after metal-on-metal THA in patients with renal failure.

Authors:  Chang Ich Hur; Taek Rim Yoon; Sang Gwon Cho; Eun Kyoo Song; Jong Keun Seon
Journal:  Clin Orthop Relat Res       Date:  2008-02-10       Impact factor: 4.176

3.  Concordance between laboratories in metal ion testing in patients with metal-on-metal hip implants.

Authors:  Raghav Saini; Pam Railton; Jessica Boyd; Hossein Sadrzadeh; James N Powell
Journal:  Can J Surg       Date:  2019-02-01       Impact factor: 2.089

4.  Hexavalent chromium induces chromosome instability in human urothelial cells.

Authors:  Sandra S Wise; Amie L Holmes; Louis Liou; Rosalyn M Adam; John Pierce Wise
Journal:  Toxicol Appl Pharmacol       Date:  2016-02-18       Impact factor: 4.219

5.  [Pathological findings in tissue surrounding revised metal/metal articulations].

Authors:  W Baur; W Hönle; H-G Willert; A Schuh
Journal:  Orthopade       Date:  2005-03       Impact factor: 1.087

6.  Immunologic adverse reaction associated with low-carbide metal-on-metal bearings in total hip arthroplasty.

Authors:  Panagiotis Aroukatos; Maria Repanti; Thomas Repantis; Vassiliki Bravou; Panagiotis Korovessis
Journal:  Clin Orthop Relat Res       Date:  2009-12-18       Impact factor: 4.176

7.  Serum Concentrations of 15 Elements Among Helicobacter Pylori-Infected Residents from Lujiang County with High Gastric Cancer Risk in Eastern China.

Authors:  Anla Hu; Li Li; Chuanlai Hu; Daoming Zhang; Chen Wang; Yan Jiang; Meng Zhang; Chunmei Liang; Wenjun Chen; Qingli Bo; Qihong Zhao
Journal:  Biol Trace Elem Res       Date:  2018-03-03       Impact factor: 3.738

8.  2008 John Charnley award: metal ion levels after metal-on-metal total hip arthroplasty: a randomized trial.

Authors:  C Anderson Engh; Steven J MacDonald; Supatra Sritulanondha; Abigail Thompson; Douglas Naudie; Charles A Engh
Journal:  Clin Orthop Relat Res       Date:  2008-10-15       Impact factor: 4.176

9.  Metal-on-Metal Total Hip Resurfacing Arthroplasty: An Evidence-Based Analysis.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2006-02-01

Review 10.  An overview of recent advances in designing orthopedic and craniofacial implants.

Authors:  Venkata P Mantripragada; Beata Lecka-Czernik; Nabil A Ebraheim; Ambalangodage C Jayasuriya
Journal:  J Biomed Mater Res A       Date:  2013-06-14       Impact factor: 4.396

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.