Literature DB >> 12556262

Evaluation of a prototypic inspiratory impedance threshold valve designed to enhance the efficiency of cardiopulmonary resuscitation.

Keith G Lurie1, Thomas A Barnes, Todd M Zielinski, Scott H McKnite.   

Abstract

OBJECTIVE: Assess a prototype inspiratory impedance threshold valve (ITV) designed to enhance vital organ circulation during standard and active compression/decompression cardiopulmonary resuscitation (CPR).
BACKGROUND: The ITV attaches to commonly used airway assist devices and decreases intrathoracic pressure during the decompression (chest recoil) phase of CPR by creating a vacuum within the thorax, which increases venous blood flow to the heart and thus increases coronary perfusion pressure and blood flow to the brain.
METHODS: The evaluation included laboratory bench testing, according to American Society for Testing and Materials (ASTM) and International Standards Organization (ISO) guidelines, and performance testing with pigs in cardiac arrest. A vacuum pull test was developed to determine the inspiratory impedance under various inspiratory flow conditions.
RESULTS: The valve passed all minimum ASTM and ISO performance tests. During cardiac arrest in pigs the ITV decreased intrathoracic pressures by 6-8 mm Hg during the decompression phase. The vacuum pull test demonstrated that the prototype ITV functioned as intended.
CONCLUSIONS: The prototype ITV passed all performance testing recommended by international guidelines and functioned effectively as intended for use. The animal study results, when combined with recent clinical data, suggest that an ITV inspiratory cracking pressure of 12 cm H(2)O should be sufficient to decrease intrathoracic pressure during the decompression phase of standard CPR. Clinical studies are now underway.

Entities:  

Mesh:

Year:  2003        PMID: 12556262

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  4 in total

Review 1.  [Treatment of hemorrhagic shock. New therapy options].

Authors:  W G Voelckel; A von Goedecke; D Fries; A C Krismer; V Wenzel; K H Lindner
Journal:  Anaesthesist       Date:  2004-12       Impact factor: 1.041

2.  Intrathoracic pressure regulation improves 24-hour survival in a pediatric porcine model of hemorrhagic shock.

Authors:  Anja Metzger; Timothy Matsuura; Scott McKnite; Bradley S Marino; Vinay M Nadkarni; Demetris Yannopoulos
Journal:  Pediatr Res       Date:  2011-09       Impact factor: 3.756

3.  Improved cerebral perfusion pressures and 24-hr neurological survival in a porcine model of cardiac arrest with active compression-decompression cardiopulmonary resuscitation and augmentation of negative intrathoracic pressure.

Authors:  Anja K Metzger; Margot Herman; Scott McKnite; Wanchun Tang; Demetris Yannopoulos
Journal:  Crit Care Med       Date:  2012-06       Impact factor: 7.598

4.  Resuscitation Outcomes Consortium (ROC) PRIMED cardiac arrest trial methods part 1: rationale and methodology for the impedance threshold device (ITD) protocol.

Authors:  Tom P Aufderheide; Peter J Kudenchuk; Jerris R Hedges; Graham Nichol; Richard E Kerber; Paul Dorian; Daniel P Davis; Ahamed H Idris; Clifton W Callaway; Scott Emerson; Ian G Stiell; Thomas E Terndrup
Journal:  Resuscitation       Date:  2008-05-19       Impact factor: 5.262

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.