Literature DB >> 12543117

A novel sensor for bite force determinations.

Cláudio P Fernandes1, Per Olof J Glantz, Stig A Svensson, Anders Bergmark.   

Abstract

OBJECTIVES: The clinical usefulness, accuracy and precision of a novel bite force sensor based on force sensing resistors were tested in six subjects wearing maxillary removable partial dentures retained by conical crowns.
METHODS: The surfaces of the sensor were manufactured in a silicone material that had mechanical properties similar to those of tough foodstuffs. In two separate series of standardized bite force tests, submaximum force levels were recorded with the sensor and with a strain gaged bite fork. Subjects were assisted in the loading tests with visual feedback instrumentation. Reliability estimates for the bite force sensor were calculated in order to show their reproducibility. Strain gages attached to the prostheses were used to determine the pattern of force distribution during loading tests. The bite force results obtained with the new bite force sensor and with the bite fork were analyzed with ANOVA and Scheffés tests. The strain patterns registered with strain gages were analyzed with F-test.
RESULTS: The bite force sensor and the bite fork transducer showed no statistically significant differences in respect of intra-individual bite force levels (range 50-300N). The bite forces registered with the new sensor were dependent on the loading position (p<0.05), sex (p<0.05) and test subject (p<0.05). The reliability of the new sensor for submaximum bite forces was calculated to be 93%. Strain gage results showed that the new sensor generated strain patterns of less variance (p<0.05) than the bite fork and therefore allowed for higher precision during biting tests. SIGNIFICANCE: The presented instrument has such clinical merits, as to favor its use in experimental clinical studies on the biomechanics of prosthetic appliances.

Entities:  

Mesh:

Year:  2003        PMID: 12543117     DOI: 10.1016/s0109-5641(02)00020-9

Source DB:  PubMed          Journal:  Dent Mater        ISSN: 0109-5641            Impact factor:   5.304


  25 in total

1.  Mastication improvement after partial implant-supported prosthesis use.

Authors:  T M S V Gonçalves; C H Campos; G M Gonçalves; M de Moraes; R C M Rodrigues Garcia
Journal:  J Dent Res       Date:  2013-10-24       Impact factor: 6.116

2.  The effect of biomechanical variables on force sensitive resistor error: Implications for calibration and improved accuracy.

Authors:  Jonathon S Schofield; Katherine R Evans; Jacqueline S Hebert; Paul D Marasco; Jason P Carey
Journal:  J Biomech       Date:  2016-02-09       Impact factor: 2.712

3.  Nano-scale sliding contact deformation behaviour of enamel under wet and dry conditions.

Authors:  Griselda Guidoni; Michael Swain; Ingomar Jäger
Journal:  J Mater Sci Mater Med       Date:  2010-02-10       Impact factor: 3.896

4.  Retention and surface changes of zirconia primary crowns with secondary crowns of different materials.

Authors:  Işıl Turp; Ergün Bozdağ; Emin Sünbüloğlu; Cem Kahruman; Ibrahim Yusufoğlu; Gülsen Bayraktar
Journal:  Clin Oral Investig       Date:  2014-01-31       Impact factor: 3.573

Review 5.  Bite Force Recording Devices - A Review.

Authors:  Tarun Prakash Verma; Kanteshwari Iranagouda Kumathalli; Vinay Jain; Rajesh Kumar
Journal:  J Clin Diagn Res       Date:  2017-09-01

6.  Comparison of wearable sensors for estimation of chewing strength.

Authors:  Delwar Hossain; Masudul Haider Imtiaz; Edward Sazonov
Journal:  IEEE Sens J       Date:  2020-01-20       Impact factor: 3.301

7.  Bite force and influential factors on bite force measurements: a literature review.

Authors:  Duygu Koc; Arife Dogan; Bulent Bek
Journal:  Eur J Dent       Date:  2010-04

8.  Occlusal force, electromyographic activity of masticatory muscles and mandibular flexure of subjects with different facial types.

Authors:  William Custodio; Simone Guimarães Farias Gomes; Fernanda Faot; Renata Cunha Matheus Rodrigues Garcia; Altair Antoninha Del Bel Cury
Journal:  J Appl Oral Sci       Date:  2011-06-03       Impact factor: 2.698

9.  Novel system for bite-force sensing and monitoring based on magnetic near field communication.

Authors:  Andres Diaz Lantada; Carlos González Bris; Pilar Lafont Morgado; Jesús Sanz Maudes
Journal:  Sensors (Basel)       Date:  2012-08-24       Impact factor: 3.576

10.  Periodontal Clinical Parameters as a Predictor of Bite Force: A Cross-Sectional Study.

Authors:  Hussain Nayef Hussain Alsharif; Kiran Kumar Ganji; Mohammad Khursheed Alam; Srinivas Munisekhar Manay; Vinod Bandela; Mohammed Ghazi Sghaireen; Mohammed Assayed Mousa
Journal:  Biomed Res Int       Date:  2021-05-10       Impact factor: 3.411

View more