Is TrueFISP a gradient-echo or a spin-echo sequence?

Klaus Scheffler1, Jürgen Hennig.   

Abstract

It is commonly accepted that TrueFISP (balanced FFE, FIESTA) belongs to the class of gradient-echo (GRE) sequences. GRE sequences are sensitive to dephasing effects of the transverse magnetization between the excitation pulse and echo acquisition, and phase coherence is only established directly after and before excitation pulses. However, an analysis of the phase evolution of transverse magnetization in a TrueFISP experiment shows very close similarities to the echo formation of a spin-echo (SE) experiment. If dephasing between excitation pulses is below +/-pi, TrueFISP exhibits a nearly complete refocusing of transverse magnetization at TE = TR/2. Only signals acquired before and after TR/2 show an additional T*2 sensitivity. Copyright 2003 Wiley-Liss, Inc.

Mesh:

Year:  2003        PMID: 12541263     DOI: 10.1002/mrm.10351

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  41 in total

1.  Centering the projection reconstruction trajectory: reducing gradient delay errors.

Authors:  Dana C Peters; J Andrew Derbyshire; Elliot R McVeigh
Journal:  Magn Reson Med       Date:  2003-07       Impact factor: 4.668

Review 2.  Principles and applications of balanced SSFP techniques.

Authors:  Klaus Scheffler; Stefan Lehnhardt
Journal:  Eur Radiol       Date:  2003-08-20       Impact factor: 5.315

3.  Incorporation of nonzero echo times in the SPGR and bSSFP signal models used in mcDESPOT.

Authors:  Mustapha Bouhrara; Richard G Spencer
Journal:  Magn Reson Med       Date:  2015-09-26       Impact factor: 4.668

4.  High sensitivity MR acoustic radiation force imaging using transition band balanced steady-state free precession.

Authors:  Yuan Zheng; Michael Marx; G Wilson Miller; Kim Butts Pauly
Journal:  Magn Reson Med       Date:  2017-06-20       Impact factor: 4.668

5.  Single heartbeat cardiac tagging for the evaluation of transient phenomena.

Authors:  Daniel A Herzka; J Andrew Derbyshire; Peter Kellman; Elliot R McVeigh
Journal:  Magn Reson Med       Date:  2005-12       Impact factor: 4.668

6.  Frequency distribution and signal formation around a vessel.

Authors:  C H Ziener; W R Bauer; P M Jakob
Journal:  MAGMA       Date:  2005-09-20       Impact factor: 2.310

7.  S5FP: spectrally selective suppression with steady state free precession.

Authors:  J A Derbyshire; D A Herzka; E R McVeigh
Journal:  Magn Reson Med       Date:  2005-10       Impact factor: 4.668

8.  Echo-dephased steady state free precession.

Authors:  Sunil Patil; Oliver Bieri; Klaus Scheffler
Journal:  MAGMA       Date:  2009-05-16       Impact factor: 2.310

9.  Improving contrast to noise ratio of resonance frequency contrast images (phase images) using balanced steady-state free precession.

Authors:  Jongho Lee; Masaki Fukunaga; Jeff H Duyn
Journal:  Neuroimage       Date:  2010-10-30       Impact factor: 6.556

10.  Self-calibrated trajectory estimation and signal correction method for robust radial imaging using GRAPPA operator gridding.

Authors:  Anagha Deshmane; Martin Blaimer; Felix Breuer; Peter Jakob; Jeffrey Duerk; Nicole Seiberlich; Mark Griswold
Journal:  Magn Reson Med       Date:  2015-03-11       Impact factor: 4.668

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.