Literature DB >> 12530830

All-solid-state calcium solvent polymeric membrane electrode for low-level concentration measurements.

Agata Michalska1, Anna Konopka, Magdalena Maj-Zurawska.   

Abstract

An all-solid-state calcium-selective electrode with a plastic membrane phase containing a calcium ionophore ETH-1001 placed on poly(3-methylthiophene) ion-to-electron transducting layer functionalized to bind calcium cations has been constructed. The obtained potentiometric sensors were characterized with a calibration line of slope close to Nernstian within the activity from 10(-5) to 0.1 M. For Ca2+ activity lower than 10(-5) M, super-Nernstian behavior was observed. The super-Nernstian response that is observed for electrodes with an internal solution strongly binding primary ions was in this case attributed to incorporation of calcium ions in the modified solid-contact phase. With this arrangement, the evaluation of selectivity coefficients much closer to those that depend on the properties of the ion-selective membrane itself was possible.

Entities:  

Year:  2003        PMID: 12530830     DOI: 10.1021/ac025916y

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  8 in total

1.  Approaches to Improving the Lower Detection Limit of Polymeric Membrane Ion-Selective Electrodes.

Authors:  Zsófia Szigeti; Tamás Vigassy; Eric Bakker; Ernö Pretsch
Journal:  Electroanalysis       Date:  2006-07-01       Impact factor: 3.223

2.  Solid-contact potentiometric polymer membrane microelectrodes for the detection of silver ions at the femtomole level.

Authors:  Nastassia Rubinova; Karin Chumbimuni-Torres; Eric Bakker
Journal:  Sens Actuators B Chem       Date:  2006-11-16       Impact factor: 7.460

Review 3.  Modern potentiometry.

Authors:  Eric Bakker; Ernö Pretsch
Journal:  Angew Chem Int Ed Engl       Date:  2007       Impact factor: 15.336

4.  Solid contact potentiometric sensors for trace level measurements.

Authors:  Karin Y Chumbimuni-Torres; Nastassia Rubinova; Aleksandar Radu; Lauro T Kubota; Eric Bakker
Journal:  Anal Chem       Date:  2006-02-15       Impact factor: 6.986

5.  Modern Directions for Potentiometric Sensors.

Authors:  Eric Bakker; Karin Chumbimuni-Torres
Journal:  J Braz Chem Soc       Date:  2008-01-01       Impact factor: 1.838

6.  Carbon-Based Solid-State Calcium Ion-Selective Microelectrode and Scanning Electrochemical Microscopy: A Quantitative Study of pH-Dependent Release of Calcium Ions from Bioactive Glass.

Authors:  Jyothir Ganesh Ummadi; Corey J Downs; Vrushali S Joshi; Jack L Ferracane; Dipankar Koley
Journal:  Anal Chem       Date:  2016-02-23       Impact factor: 6.986

7.  Effects of architecture and surface chemistry of three-dimensionally ordered macroporous carbon solid contacts on performance of ion-selective electrodes.

Authors:  Melissa A Fierke; Chun-Ze Lai; Philippe Bühlmann; Andreas Stein
Journal:  Anal Chem       Date:  2010-01-15       Impact factor: 6.986

8.  Subnanomolar Detection Limit Application of Ion-Selective Electrodes with Three-Dimensionally Ordered Macroporous (3DOM) Carbon Solid Contacts.

Authors:  Chun-Ze Lai; Marti M Joyer; Melissa A Fierke; Nicholas D Petkovich; Andreas Stein; Philippe Bühlmann
Journal:  J Solid State Electrochem       Date:  2009-01-01       Impact factor: 2.647